Correlación de gasto cardiaco obtenido por termodilución y por monitor ultrasónico (USCOM) en pacientes con choque séptico / Correlation of cardiac output obtained by termodulution and by ultrasonic monitor in patients with septic shock (USCOM) / Correlação do débito cardíaco obtido pela termodulação e pelo monitor ultrasônico (USCOM) em pacientes com choque septico
Cruz Jiménez, María del Carmen; Carmona Domínguez, Aurea; Tapia Velasco, Rafael; Nava López, Jorge Arturo; Montelongo, Felipe de Jesús.
  • Cruz Jiménez, María del Carmen; Instituto de Salud del Estado de México Las Américas. Ecatepec. MX
  • Carmona Domínguez, Aurea; Instituto de Salud del Estado de México Las Américas. Ecatepec. MX
  • Tapia Velasco, Rafael; Instituto de Salud del Estado de México Las Américas. Ecatepec. MX
  • Nava López, Jorge Arturo; Instituto de Salud del Estado de México Las Américas. Ecatepec. MX
  • Montelongo, Felipe de Jesús; Instituto de Salud del Estado de México Las Américas. Ecatepec. MX
Med. crít. (Col. Mex. Med. Crít.) ; 32(4): 191-200, jul.-ago. 2018. tab, graf
Article in Spanish | LILACS-Express | LILACS | ID: biblio-1114981
RESUMEN
Resumen El catéter en la arteria pulmonar (CAP) es un dispositivo utilizado en unidades de cuidados intensivos (UCI) para medir las presiones en el corazón y los vasos sanguíneos pulmonares como parte del monitoreo hemodinámico, principalmente en pacientes de cirugía cardiaca. El dispositivo USCOM se trata de una técnica no invasiva que utiliza la tecnología Doppler para obtener las medidas de volumen sistólico y sus derivados. Se realiza la siguiente comparación de medición de GC entre estos dos dispositivos en pacientes con choque séptico. Se realizó un estudio tipo observacional, prospectivo, longitudinal y comparativo en pacientes con choque séptico entre 18 y 60 años de edad ingresados en la UTI en el periodo de mayo-junio del 2017. Ante la disminución del uso del catéter de la arteria pulmonar debido a la controversia de no mejorar la mortalidad en los pacientes de las unidades de terapia intensiva (UTI), la colocación de dicho catéter ha caído en desuso; sin embargo, el GC medido por el catéter de Swan-Ganz sigue siendo el «estándar de oro¼ para la medición en tiempo real del GC y las resistencias sistémicas y pulmonares. La medición del GC por CAP versus USCOM se correlaciona de tal forma que puede emplearse en la medición por USCOM en un paciente con choque séptico, al cual no se le pretenda invadir para determinar sus condiciones hemodinámicas.
ABSTRACT
Abstract The pulmonary artery catheter (CAP) is a device used in intensive care units (ICUs) to measure pressures in the heart and pulmonary blood vessels as part of hemodynamic monitoring primarily in cardiac surgery patients. The USCOM device is a non-invasive technique that uses Doppler technology to obtain measurements of systolic volume and its derivatives. The following CO measurement comparison is performed between these two devices in patients with septic shock. An observational, prospective, longitudinal and comparative study was conducted in patients with septic shock aged between 18 and 60 years admitted to intensive care in the period May-June 2017. In view of the decrease in the use of the pulmonary artery catheter due to the controversy of not improving the mortality in the patients of the Intensive Care Units, the placement of this catheter has fallen into disuse; however, cardiac output measured by the Swan Ganz catheter remains the «gold standard¼ for real-time measurement of cardiac output and systemic and pulmonary resistance. The CO measurement by PAC versus USCOM correlates, in such a way, that USCOM measurement can be used in a patient with septic shock, who is not expected to invade to determine their hemodynamic conditions.
RESUMO
Resumo O cateter de artéria pulmonar (CAP) é um dispositivo utilizado em unidades de terapia intensiva (UTI) para medir as pressões nos vasos sanguíneos cardíacos e pulmonares, como parte da monitorização hemodinâmica, principalmente em pacientes submetidos a cirurgia cardíaca. O dispositivo USCOM é uma técnica não invasiva que utiliza a tecnologia Doppler para obter medidas do volume sistólico e seus derivados. A seguinte comparação da medição do DC é feita entre esses dois dispositivos em pacientes com choque séptico. Foi realizado um estudo observacional, prospectivo, longitudinal e comparativo em pacientes com choque séptico com idade entre 18 e 60 anos internados na unidade de terapia intensiva no período de maio a junho de 2017. Dada a diminuição do uso do cateter de artéria pulmonar devido à controvérsia de não melhorar a mortalidade nos pacientes das Unidades de Terapia Intensiva, a colocação do referido cateter caiu em desuso; no entanto, o débito cardíaco medido pelo cateter de Swan Ganz continua sendo o «padrão ouro¼ para a medição em tempo real do débito cardíaco e resistências sistêmicas e pulmonares. A medida do DC por CAP vs USCOM está correlacionada, de tal forma que a medida por USCOM pode ser usada em um paciente com choque séptico, que não se destina a invadir para determinar suas condições hemodinâmicas.

Full text: Available Index: LILACS (Americas) Type of study: Health economic evaluation / Observational study Language: Spanish Journal: Med. crít. (Col. Mex. Med. Crít.) Journal subject: Cirurgia Geral / Oftalmologia / Ortopedia Year: 2018 Type: Article Affiliation country: Mexico Institution/Affiliation country: Instituto de Salud del Estado de México Las Américas/MX

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Type of study: Health economic evaluation / Observational study Language: Spanish Journal: Med. crít. (Col. Mex. Med. Crít.) Journal subject: Cirurgia Geral / Oftalmologia / Ortopedia Year: 2018 Type: Article Affiliation country: Mexico Institution/Affiliation country: Instituto de Salud del Estado de México Las Américas/MX