Your browser doesn't support javascript.
loading
Analise dos dados biométricos gerados por interferometria comparada com Scheimpflug / Analysis of biometric data generated by interferometry compared with Scheimpflug
Rodrigues, Francisco Wellington; Freitas, Maria Paula Bessa de; Oliveira, Juliane Nogueira de; Silva, Rodrigo Egídio da; Chater, Salim Bosco; Chater Filho, Jamil.
  • Rodrigues, Francisco Wellington; Hospital VER. Goiânia. BR
  • Freitas, Maria Paula Bessa de; Pontifícia Universidade Católica de Goiás. Curso Acadêmico de Medicina. Goiânia. BR
  • Oliveira, Juliane Nogueira de; Pontifícia Universidade Católica de Goiás. Curso Acadêmico de Medicina. Goiânia. BR
  • Silva, Rodrigo Egídio da; Hospital VER. Goiânia. BR
  • Chater, Salim Bosco; Centro Integrado de Oftalmologia San Charbel. Goiânia. BR
  • Chater Filho, Jamil; Centro Integrado de Oftalmologia San Charbel. Goiânia. BR
Rev. bras. oftalmol ; 79(5): 289-293, set.-out. 2020. tab
Article in Portuguese | LILACS | ID: biblio-1137987
RESUMO
Resumo

Objetivo:

Observar o grau de concordância das variáveis analisadas entre os dispositivos IOL Master 500 e Pentacam AXL e descrever as medias

Métodos:

Foram analisados 35 prontuários, totalizando 61 olhos. Todos os pacientes se submeteram à avaliação biométrica nos dois dispositivos, no período de agosto de 2018 a agosto de 2019. Os dados coletados foram idade, sexo, profundidade da câmara anterior, comprimento axial, K1, K2, poder dióptrico da LIO e alvo refracional.

Resultados:

As médias das variáveis analisadas entre os dispositivos de biométricos óptica em questão tiveram diferença estatisticamente significante (p<0,05). A regressão linear não apontou influência de nenhuma das variáveis da câmara anterior na diferença de valores do poder dióptrico da LIO e do alvo refracional entre os dispositivos.

Conclusão:

Não houve concordância estatística entre os dispositivos para as variáveis analisadas. Portanto, deve se evitar intercambiar o uso do Pentacam AXL com o IOL Master 500.
ABSTRACT
Abstract

Objective:

Observe the agreement between IOL Master 500 and Pentacam AXL and describe the averages.

Methods:

We analyzed 35 medical records, totaling 61 eyes. All patients underwent biometric evaluation on both devices from August 2018 to August 2019. The data collected were age, gender, anterior chamber depth, axial length, K1, K2, biometrics and IOL target.

Results:

The averages of the variables analyzed between the optical biometric devices in question had a statistically significant difference (p <0.05). Linear regression showed no influence of any anterior chamber variables on the difference in biometrics and target values between the devices.

Conclusion:

There was no statistical agreement between the devices for the analyzed variables. Therefore, the interchange of Pentacam AXL with IOL Master 500 should be avoided.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Refraction, Ocular / Biometry / Axial Length, Eye / Multifocal Intraocular Lenses / Interferometry Limits: Aged / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2020 Type: Article Affiliation country: Brazil Institution/Affiliation country: Centro Integrado de Oftalmologia San Charbel/BR / Hospital VER/BR / Pontifícia Universidade Católica de Goiás/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Refraction, Ocular / Biometry / Axial Length, Eye / Multifocal Intraocular Lenses / Interferometry Limits: Aged / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2020 Type: Article Affiliation country: Brazil Institution/Affiliation country: Centro Integrado de Oftalmologia San Charbel/BR / Hospital VER/BR / Pontifícia Universidade Católica de Goiás/BR