Identificación directa de microorganismos a partir de hemocultivos positivos utilizando espectrometría de masas (MALDI-TOF MS) / Direct identification of microorganisms in positive blood culture bottles by mass spectrometry (MALDI-TOF MS)
Azula, Natalia; Relloso, Silvia; Barrios, Rubén; Nicola, Federico; Smayevsky, Jorgelina.
  • Azula, Natalia; Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC. AR
  • Relloso, Silvia; Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC. AR
  • Barrios, Rubén; BD Diagnostics Systems. AR
  • Nicola, Federico; Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC. AR
  • Smayevsky, Jorgelina; Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC. AR
Acta bioquím. clín. latinoam ; 56(1): 17-31, ene. 2022. graf
Article in Spanish | LILACS, BINACIS | ID: biblio-1402943
RESUMEN
Resumen La espectrometría de masas (MALDI-TOF MS) permite la identificación de microorganismos directamente de las colonias en pocos minutos. En este estudio se ha desarrollado y evaluado un protocolo reducido para identificar microorganismos directamente de las botellas de hemocultivos positivos en 30 minutos con una alta sensibilidad y especificidad, utilizando MALDITOF. Un total de 2535 hemocultivos positivos fueron estudiados por el método directo de MALDI-TOF MS, a partir de una alícuota de sangre de las botellas y el método de colonia, utilizando los cultivos desarrollados en medios sólidos. Del total de hemocultivos positivos incluidos en este estudio, 2381 (93,9%) fueron monomicrobianos y 146 (5,8%) polimicrobianos. Mil trescientos treinta (55,9%) de los aislamientos correspondieron a cocos gram positivos, 922 (38,7%) a bacilos gram negativos, 60 (2,5%) a anaerobios, 36 (1,5%) a bacilos gram positivos y 13 a levaduras. La concordancia global entre ambos métodos fue del 81,7% a nivel de especie (90,0% para bacilos gram negativos, 76,7% para cocos gram positivos y 33,3% para bacilos gram positivos). Se identificó al menos un germen en el 88% de las botellas positivas con desarrollo polimicrobiano. Los resultados del presente estudio demostraron que el protocolo basado en MALDI-TOF MS permite la identificación microbiana directamente de hemocultivos positivos en un tiempo corto, con una alta precisión, con excepción de los bacilos gram positivos.
ABSTRACT
Abstract Matrix-assisted laser desorption ionization-time of flight mass spectrometry (MALDI-TOF MS) enables the identification of microorganisms directly from colonies within minutes. In this study this technology was adapted and tested for use with blood culture bottles, thus allowing identification in 30 minutes once the blood culture is detected as positive by the automate. A total of 2535 blood culture bottles reported as positive were tested by MALDI-TOF MS directly from positive blood culture bottles and colonies. A total of 2381 (93.9%) and 146 (5.8%) of the positive blood cultures were monomicrobial and polymicrobial, respectively. And 1330 (55.9%), 922 (38.7%), 60 (2.5%), 36 (1.5%) and 13 of the isolates were gram-positive cocci (GPC), gram-negative bacilli (GNB), anaerobic bacteria, gram-positive bacilli (GPB) and yeast respectively. Concordance between both methods was 81.7% (76.7% of GPC, 90% of GNB, 74.2% of anaerobic bacteria and 33.3% of GPB) in monomicrobial cultures. Eighty eight per cent of the polymicrobial cultures were identified correctly in at least one of the two bacteria. The results of the present study show that this fast, MALDI-TOF MS based method allows microbial identification directly from positive blood culture in a short time, with a high accuracy, with the exception of gram-positive bacilli.
RESUMO
Resumo A espectrometria de massa (MALDI-TOF MS) permite a identificação de microorganismos diretamente das colônias em minutos. Nesse estudo, foi desenvolvido um protocolo reduzido para identificar microrganismos diretamente das garrafas de hemoculturas positivas em 30 minutos com alta sensibilidade e especificidade, utilizando MALDI-TOF. Um total de 2535 hemoculturas positivas foram relatadas -o método direto de MALDI-TOF MS, a partir de uma alíquota de sangue dos vidros e o método de colônia, a partir das culturas desenvolvidas em meios sólidos. Do total de hemoculturas positivas incluídas neste estudo, 2.381 (93,9%) eram monomicrobianas e 146 (5,8%) eram polimicrobianas. Mil trezentos e trinta (55,9%) dos isolados corresponderam a cocos gram-positivos, 922 (38,7%) bacilos gram-negativos, 60 (2,5%) anaeróbios, 36 (1,5%) bacilos gram-positivos e 13 leveduras. A concordância geral entre os dois métodos foi de 81,7% em nivel de especie (90,0% para bacilos gram-negativos, 76,7% para cocos gram-positivos e 33,3% para bacilos gram-positivos). Pelo menos um germe foi identificado em 88% dos vidros positivos com desenvolvimento polimicrobiano. Os resultados do presente estudo demonstraram que o protocolo baseado em MALDI-TOF MS permite a identificação microbiana diretamente de hemoculturas positivas em um curto espaço de tempo, com alta precisão, com exceção de bacilos gram-positivos.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Mass Spectrometry / Gram-Positive Rods / Microbiology Type of study: Diagnostic study Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2022 Type: Article Affiliation country: Argentina Institution/Affiliation country: BD Diagnostics Systems/AR / Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC/AR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Mass Spectrometry / Gram-Positive Rods / Microbiology Type of study: Diagnostic study Language: Spanish Journal: Acta bioquím. clín. latinoam Journal subject: Bioqu¡mica / Qu¡mica Cl¡nica Year: 2022 Type: Article Affiliation country: Argentina Institution/Affiliation country: BD Diagnostics Systems/AR / Centro de Educación Médica e Investigaciones Clínicas Norberto Quirno - CEMIC/AR