Aplicación de un factor de corrección para el cálculo del lente intraocular en pacientes con cámara anterior / Application of a correction factor for intraocular lens power calculation in patients with a narrow anterior chamber and augmented crystalline lens thickness
Miranda Hernández, Iramis; Gutiérrez Castillo, Marietta; de Almeida, Yokana; Barroso Lorenzo, Raúl; Ramos Pereira, Yanay; Sibila González, Marlene E.
  • Miranda Hernández, Iramis; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Gutiérrez Castillo, Marietta; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • de Almeida, Yokana; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Barroso Lorenzo, Raúl; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Ramos Pereira, Yanay; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
  • Sibila González, Marlene E; Instituto Cubano de Oftalmología "Ramón Pando Ferrer". La Habana. CU
Rev. cuba. oftalmol ; 31(2)abr.-jun. 2018.
Article in Spanish | CUMED, LILACS | ID: biblio-1508346
RESUMEN

Objetivo:

evaluar la aplicación de un factor corrección en el cálculo del lente intraocular en pacientes con cámara anterior estrecha y grosor del cristalino aumentado.

Métodos:

se realizó un estudio experimental donde se aplicó un factor de corrección y se comparó con dos grupos de control.

Resultados:

predominó el sexo femenino con el 71 por ciento de los casos. La edad fue de 70 años y más. El 48,4 por ciento de los pacientes presentó una esfera posoperatoria entre ± 0,50 dioptrías; el 19,4 por ciento estuvo por debajo de -0,50 dioptrías y el 32,3 por ciento de los pacientes por encima de + 0,50 dioptrías. El grupo 3 (grosor del cristalino mayor que 4,60 mm si factor de corrección) tuvo el mayor porcentaje de esfera posoperatoria ± 0,50 dioptrías (58,3 por ciento). El grupo 2 tuvo el mayor porcentaje de pacientes con esfera obtenida mayor de 0,50 dioptrías (38,2 por ciento).

Conclusiones:

los pacientes a quienes se les aplica el factor de corrección obtienen una esfera posoperatoria cercana a la emetropía a pesar de que la muestra no es homogénea y no se obtienen esferas por encima de 1 dioptría(AU)
ABSTRACT

Objective:

Evaluate the application of a correction factor for intraocular lens power calculation in patients with a narrow anterior chamber and augmented crystalline lens thickness.

Methods:

An experimental study was conducted in which a correction factor was applied and compared with two control groups.

Results:

Female sex prevailed with 71 percent of the cases. Age was 70 years and over. 48.4 percent of the patients had a postoperative sphere between ± 0.50 diopters; 19.4 percent were below - 0.50 diopters and 32.3 percent were above + 0.50 diopters. Group 3 (crystalline lens thickness above 4.60 mm without correction factor) had the highest postoperative sphere percentage of ± 0.50 diopters (58.3 percent). Group 2 had the highest percentage of patients with an achieved sphere above 0.50 diopters (38.2 percent).

Conclusions:

Patients to whom the correction factor was applied achieved a postoperative sphere close to emmetropia, despite the fact that the sample was not homogeneous and spheres above 1 diopter were not obtained(AU)
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Lenses, Intraocular Limits: Aged / Female / Humans Language: Spanish Journal: Rev. cuba. oftalmol Journal subject: Ophthalmology Year: 2018 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto Cubano de Oftalmología "Ramón Pando Ferrer"/CU

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Lenses, Intraocular Limits: Aged / Female / Humans Language: Spanish Journal: Rev. cuba. oftalmol Journal subject: Ophthalmology Year: 2018 Type: Article Affiliation country: Cuba Institution/Affiliation country: Instituto Cubano de Oftalmología "Ramón Pando Ferrer"/CU