Objetivo:

avaliar o torque de inserção de um novo desenho de implante. Material e

métodos:

pacientes foram recrutados para instalação de implantes, no período de março a dezembro de 2015. Dados como o diâmetro, o comprimento do implante, a sequência de fresagem, a área de instalação e o torque atingido foram considerados. A avaliação do torque foi realizada com um torquímetro cirúrgico.

Resultados:

foram avaliados 82 implantes, com diâmetros entre 3,5 mm e 5 mm, e comprimentos entre 7 mm e 16 mm, sendo instalados 46 implantes na mandíbula e 36 implantes na maxila. Os torques (média ± desvio-padrão) alcançados foram de 53 ± 9,9 Ncm para maxila e 57 ± 11,3 Ncm para mandíbula.

Conclusão:

o sistema de implantes Novo Colosso promoveu torque de inserção adequado, com boa estabilidade primária nos implantes.
ABSTRACT

Objective:

to evaluate the insertion torque of a new implant design. Material and

methods:

patients were recruited for implant placement from March to December 2015. Data such as diameter, implant length, drilling sequence, installation area, and torque reached were considered. Torque evaluation was performed with a surgical torque wrench.

Results:

82 implants were evaluated, with diameters between 3.5 mm and 5.0 mm, and lengths between 7 mm and 16 mm, with 46 implants in the mandible and 36 implants in the maxilla. The torques (mean ± standard deviation) reached 53 ± 9.9 Ncm for maxilla and 57 ± 11.3 Ncm for mandible.

Conclusion:

the Novo Colosso implant system promotes adequate insertion torque with good primary implant stability.
Subject(s)

Search on Google
Index: LILACS (Americas) Main subject: Osteotomy / Dental Implants / Osseointegration / Torque Type of study: Evaluation studies Limits: Humans Language: Portuguese Journal: ImplantNewsPerio Journal subject: Odontologia Year: 2017 Type: Article Affiliation country: Brazil Institution/Affiliation country: Unesp/BR

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: LILACS (Americas) Main subject: Osteotomy / Dental Implants / Osseointegration / Torque Type of study: Evaluation studies Limits: Humans Language: Portuguese Journal: ImplantNewsPerio Journal subject: Odontologia Year: 2017 Type: Article Affiliation country: Brazil Institution/Affiliation country: Unesp/BR