Comparação entre Técnicas de Block-Matching e Optical Flow na análise do Strain pela Ecocardiografia / Comparison of Block-Matching and Optical Flow in Strain Echocardiography Analysis
Linhares, Renata Rejane; Barreto, Rodrigo Bellio de Mattos; Silva, Carlos Eduardo Suaíde; Le Bihan, David Costa de Souza; Peixoto, Luciana Braz; Cruz, Andresa Paes da; Garcia, Luara Piovan; Monaco, Claudia Gianini; Gil, Manuel Adán.
  • Linhares, Renata Rejane; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Barreto, Rodrigo Bellio de Mattos; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Silva, Carlos Eduardo Suaíde; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Le Bihan, David Costa de Souza; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Peixoto, Luciana Braz; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Cruz, Andresa Paes da; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Garcia, Luara Piovan; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Monaco, Claudia Gianini; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
  • Gil, Manuel Adán; DASA - Diagnóstico das Américas SA. São Paulo - SP. BR
ABC., imagem cardiovasc ; 30(3): f:87-l:91, jul.-set. 2017. tab, ilus
Article in Portuguese | LILACS | ID: biblio-848719
RESUMO

Introdução:

Diferentes tecnologias têm sido utilizadas para avaliar a função sistólica do ventrículo esquerdo. De particular interesse, está a ecocardiografia com strain bidimensional (2DSTE). Dois métodos diferentes têm sido usados para quantificar a deformação miocárdica (strain miocárdico) pela 2DSTE block matching e optical flow. Ambos estão presentes em ecocardiógrafos comercialmente disponíveis. Entretanto, não há consenso a respeito se as medidas do strain longitudinal por estes métodos são sobreponíveis.

Objetivo:

Comparar os valores de pico do strain miocárdico longitudinal obtidos através da 2DSTE pelos dois diferentes métodos (block matching x optical flow).

Método:

Foram realizadas as medidas do strain em 16 segmentos miocárdicos do ventrículo esquerdo, conforme orientação da Sociedade Americana de Ecocardiografia pela técnica do block matching (Vivid 7, GE, Horten, Noruega) e, imediatamente após, pela técnica do optical flow (My Lab 60, Esaote, Florença, Itália) de forma randomizada e seus valores foram comparados.

Resultados:

Houve 28 indivíduos com idade 27,9 ± 7,7 anos, sendo 50% do sexo masculino e todos com fração de ejeção superior a 55%. A medida do strain longitudinal global foi maior pela técnica de block matching (p = 0,02). Na análise dos 16 segmentos, houve 10 segmentos com valores diferentes, em especial os segmentos apicais.

Conclusão:

Os valores do strain miocárdico longitudinal obtidos por métodos diferentes não são sobreponíveis e devem ser usados com cautela. Sendo assim, valores de normalidade também variam de acordo com o fabricante e o método utilizado
ABSTRACT

Introduction:

Different technologies have been used to evaluate left ventricular systolic function. Of particular interest is the two-dimensional strain echocardiography (2DSTE). Two different methods have been used to quantify myocardial strain by 2DSTE block matching and optical flow. Both are present in commercially available echocardiographs. However, there is no consensus as to whether the longitudinal strain measures using these methods are overlapping.

Objective:

To compare the longitudinal myocardial strain peak values obtained through 2DSTE using two different methods (block matching x optical flow).

Method:

Strain measurements in 16 left ventricular myocardial segments were taken as recommended by the American Society of Echocardiography using block matching (Vivid 7, GE, Horten, Norway) and, immediately after, using optical flow (My Lab 60, Esaote, Florence, Italy) in a randomized way and their values were compared.

Results:

There were 28 individuals aged 27.9 ± 7.7, of whom 50% were male and all with an ejection fraction greater than 55%. Global longitudinal strain was higher using block matching (p = 0.02). In the analysis of the 16 segments, 10 segments had different values, especially the apical segments.

Conclusion:

Longitudinal myocardial strain values obtained by different methods are not overlapping and must be used with caution. Normality values also vary according to the manufacturer and the method used
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Image Processing, Computer-Assisted / Diagnostic Imaging / Echocardiography, Doppler Type of study: Controlled clinical trial / Diagnostic study / Etiology study / Risk factors Limits: Adult / Female / Humans / Male Language: Portuguese Journal: ABC., imagem cardiovasc Journal subject: Cardiology / Diagnostic Imaging Year: 2017 Type: Article Affiliation country: Brazil Institution/Affiliation country: DASA - Diagnóstico das Américas SA/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Image Processing, Computer-Assisted / Diagnostic Imaging / Echocardiography, Doppler Type of study: Controlled clinical trial / Diagnostic study / Etiology study / Risk factors Limits: Adult / Female / Humans / Male Language: Portuguese Journal: ABC., imagem cardiovasc Journal subject: Cardiology / Diagnostic Imaging Year: 2017 Type: Article Affiliation country: Brazil Institution/Affiliation country: DASA - Diagnóstico das Américas SA/BR