Your browser doesn't support javascript.
loading
Estudo de comparação por extensometria entre dois modelos de implante: extraoral e intraoral (estudo piloto) / Comparison study for extensometry between two implant models: extra-oral and intraoral (pilot study)
Barabarán, Pedro Miguel; Hung, Celeste Cecília Urdaniga; Gamarra, Rodrigo Ernesto Salazar; Moura, Renata de Vasconcellos; Dib, Luciano Lauria; Mesquita, Alfredo Mikail Melo.
  • Barabarán, Pedro Miguel; Universidade Paulista. São Paulo. BR
  • Hung, Celeste Cecília Urdaniga; Universidade Paulista. São Paulo. BR
  • Gamarra, Rodrigo Ernesto Salazar; Universidade Paulista. São Paulo. BR
  • Moura, Renata de Vasconcellos; Universidade Paulista. São Paulo. BR
  • Dib, Luciano Lauria; Universidade Paulista. São Paulo. BR
  • Mesquita, Alfredo Mikail Melo; Universidade Paulista. São Paulo. BR
ImplantNewsPerio ; 3(1): 45-50, jan.-fev. 2018. ilus
Article in Portuguese | LILACS | ID: biblio-881510
RESUMO
Objetivo: avaliar, por meio de extensometria, a infl uência da flange do implante de titânio na deformação superficial do osso. Material e métodos: um bloco de poliuretano simulando um osso tipo I mais corticalizado e menos medular (40 PCF) recebeu um implante de hexágono externo intraoral, sem flange 3,75 mm x 8,5 mm (G1) e um implante de hexágono externo extraoral, com flange 3,75 mm x 8 mm (G2). Em seguida, foram instalados componentes standard com 2 mm de altura de cinta metálica e cilindros metálicos para restauração provisória. Dois extensômetros lineares foram colados na superfície do bloco (mesial e distal de cada implante), e uma carga axial de 100 Ncm em uma máquina de ensaio universal foi aplicada, com cinco medições de microdeformação (µÎµ) para cada implante. Os dados foram submetidos ao teste Shapiro-Wilk para verifi car a homogenidade dos dados e em seguida ao teste t Student. Resultados: as médias e desvios-padrão obtidos foram: G1: 148,2 ± 5,84 e G2: 92,4 ± 11,24 µÆ, com diferença estatística significativa entre os grupos. Conclusão: os implantes extraorais tiveram melhor comportamento mecânico para as cargas axiais do que os implantes intraorais, sendo que a flange melhorou a distribuição da carga na superfície do bloco de poliuretano.
ABSTRACT
Objectives: to evaluate, by means of extensometry, the influence of the titanium implant flange on the superficial deformation of the bone. Material and methods: a polyurethane block simulating a more corticalized and less spinal type I bone (40 PCF) received an external intraoral hexagon implant with a 3.75 mm x 8.5 mm fl ange (G1) and an extraoral external hexagon implant, with flange 3.75 mm x 8 mm (G2). Subsequently, standard components with 2 mm high of metal strap and metal cylinders were installed for temporary restoration. Two linear strain gauges were glued to the surface of the block (mesial and distal of each implant), and an axial load of 100N in a universal test machine was applied with five microdemorphometric measurements (µÎµ) for each implant. The data were submitted to the Shapiro-Wilks test to verify the homogeneity of the data and then to the t-student test. Results: the means and standard deviations obtained were: G1: 148.2 ± 5.84 and G2: 92.4 ± 11.24 µÎµ, with signifi cant statistical difference between the groups. Conclusion: extraoral implants had better mechanical behavior for axial loads than intraoral implants, and the flange improved the load distribution on the surface of the polyurethane block.
Subject(s)

Search on Google
Index: LILACS (Americas) Main subject: Prostheses and Implants / Comparative Study / Osseointegration / Maxillofacial Prosthesis Implantation / Maxillofacial Prosthesis Limits: Humans Language: Portuguese Journal: ImplantNewsPerio Journal subject: Odontologia Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Paulista/BR

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: LILACS (Americas) Main subject: Prostheses and Implants / Comparative Study / Osseointegration / Maxillofacial Prosthesis Implantation / Maxillofacial Prosthesis Limits: Humans Language: Portuguese Journal: ImplantNewsPerio Journal subject: Odontologia Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Paulista/BR