Comportamento da pressão intraocular no tonômetro de aplanação de Goldman e pneumático durante o teste de sobrecarga hídrica / Intraocular pressure behavior in the Goldmann and pneumatic tonometer during water drinking test
Silva, Rodrigo Egidio da; Paiva, Isabella Cristina de Sousa; Souza, Juliana Ferreira de; Rodrigues, Francisco Wellington.
  • Silva, Rodrigo Egidio da; Pontifícia Universidade Católica de Goiás. Goiânia. BR
  • Paiva, Isabella Cristina de Sousa; Pontifícia Universidade Católica de Goiás. Goiânia. BR
  • Souza, Juliana Ferreira de; Pontifícia Universidade Católica de Goiás. Goiânia. BR
  • Rodrigues, Francisco Wellington; Pontifícia Universidade Católica de Goiás. Goiânia. BR
Rev. bras. oftalmol ; 77(1): 14-19, jan.-fev. 2018. tab, graf
Article in Portuguese | LILACS | ID: biblio-899111
RESUMO
Resumo

Objetivo:

Avaliar a concordância entre os tonômetros de aplanação de Goldman e pneumático na realização do teste de sobrecarga hídrica (TSH).

Métodos:

Estudo descritivo transversal composto por uma amostra de 102 olhos proveniente de um hospital particular em Goiânia (GO) entre 2013 a 2016, com avaliação das diferentes pressões intraoculares (PIO) nos tonômetros de aplanação de Goldman e pneumático quando submetidos ao TSH.

Resultados:

A média de idade foi de 52,17 (± 15,21) anos, sendo que60,8% dos pacientes pertenciam ao sexo feminino e 39,2% ao masculino. A média da espessura da córnea foi de 531,9(± 72,75) micra. Por regressão linear as variaveis idade e espessura da cornea central não ocorreram significancia estatistica entre os dois dispositivos analisados.

Conclusão:

Observou-se boa concordância nas medições entre os aparelhos de aplanação e o pneumático durante o teste de sobrecarga hídrica, porém necessita-se de novos estudos de maior impacto epidemiológico para confirmação desta assertiva.
ABSTRACT
Abstract

Objective:

To evaluate the consonance betweenthe Goldman and pneumática planation to nometers under the effect of the water drink test.

Methods:

Cross-sectional descriptive study consisting of a sample of 102 eyes from a private hospital in Goiânia (GO) from 2013 to 2015,with na evaluation of different intraocular pressures (IOP) in the Goldman and pneumatic flattening to nometers when submitted to TSH.

Results:

The average age was 52.17 (± 15.21) years old, 60.8% of the patients were female and 39.2% were male.The mean corneal thickness was 531.9(± 72.75) By linear regression the variables age and corneal thickness did not occur statistical significance between the two devices analyzed.

Conclusion:

Good agreement was observed in the measurements between the applanation devices and the tire during the water drink test, but new studies with a greater epidemiological impact were required to confirm this assertion.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Tonometry, Ocular / Drinking Water / Glaucoma / Ocular Hypertension / Drinking / Intraocular Pressure Type of study: Observational study / Prevalence study / Risk factors Limits: Female / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Pontifícia Universidade Católica de Goiás/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Tonometry, Ocular / Drinking Water / Glaucoma / Ocular Hypertension / Drinking / Intraocular Pressure Type of study: Observational study / Prevalence study / Risk factors Limits: Female / Humans / Male Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Pontifícia Universidade Católica de Goiás/BR