Análise da diferença entre os limiares auditivos comportamentais e os resultados obtidos através do exame potencial evocado auditivo de estado estável (PEAEE) em adultos jovens / Analysis of difference between the behavioral hearing thresholds and the results obtained through the auditory steady state response (ASSR) in young adults / Análisis de la diferencia entre los umbrales auditivos comportamentales y los resultados obtenidos a través del examen potencial evocado auditivo de estado estable (PEAEE) en adultos jóvenes
Lamenza, Alessandra Rabello de Oliveira; Castro, Jair de Carvalho e; Oliveira, Daniela Cecílio Capra Marques de.
  • Lamenza, Alessandra Rabello de Oliveira; s.af
  • Castro, Jair de Carvalho e; Universidade Federal do Rio de Janeiro. Rio de Janeiro. BR
  • Oliveira, Daniela Cecílio Capra Marques de; s.af
Distúrb. comun ; 30(2): 376-384, jun. 2018. ilus, tab
Article in Portuguese | LILACS | ID: biblio-911151
RESUMO

Introdução:

A audição, sentido que nos permite receber a informação sonora, se apresenta como um dos facilitadores para uma interação social eficaz. Um déficit nessa função acarreta prejuízo no processo comunicativo, devendo ser diagnosticado e tratado precocemente. Atualmente dispomos de meios subjetivos e objetivos para avaliar o sistema auditivo. A audiometria tonal é a forma subjetiva mais aplicada para medir esse sentido. Entretanto, em populações impossibilitadas de responder subjetivamente, aplicam-se técnicas objetivas, como o Potencial Evocado Auditivo de Estado Estável. Ambos pesquisam respostas auditivas do sujeito sob teste, utilizando técnicas distintas.

Objetivo:

Avaliar adultos jovens, sem queixas auditivas, pesquisando a sensibilidade auditiva e, a partir de indicadores estatísticos, calcular as médias das diferenças das respostas obtidas nesses procedimentos, verificando se há correlação entre eles.

Método:

Pesquisa de respostas auditivas via Audiometria Tonal e Potencial Evocado Auditivo de Estado Estável, nas frequências de 500, 1000, 2000 e 4000 Hz, em 30 sujeitos normo-ouvintes.

Resultados:

Ao calcularmos a média das diferenças entre as frequências testadas nos dois procedimentos, elas variaram de 10,47 a 18,22, não havendo forte correlação em nenhuma delas, deixando dúvidas na frequência de 1000 Hz.

Conclusão:

Podemos dizer que as médias das diferenças entre os valores obtidos nos dois procedimentos foram razoavelmente elevadas, principalmente em 500 Hz e, em menor proporção, para 4 kHz. Os valores obtidos nos dois exames, embora não tenham apresentado forte correlação, apresentaram-se discretamente melhores para 500 Hz. Resultados de 1000 Hz não nos permitem afirmar, inclusive, se existe alguma correlação entre os testes.
ABSTRACT

Introduction:

Hearing is the sense which allows us to receive auditory information; therefore it facilitates efficient social interaction. Any loss in this function damages the communicative process, so it must be diagnosed and treated as soon as possible. Nowadays, both objective and subjective hearing tests are available. Pure Tone Audiometry is the most widely used subjective way to evaluate this sense. However, for those who are unable to respond subjectively, we can try objective techniques like the Auditory Steady State Response. Both will search for patients' minimum hearing responses, applying different approaches.

Objective:

To evaluate normal-hearing young adults with no hearing complaints by analyzing their hearing sensibility and, based on statistical indicators, calculate the average of the differences between the results obtained in both procedures in order to verify the existence of correlation between them.

Method:

We studied the hearing responses from 30 normal-hearing subjects through Pure Tone Audiometry and Auditory Steady State Response at 500, 1000, 2000 and 4000 Hz.

Results:

When we calculated the average of the differences between the tested frequencies, they ranged from 10,47 to 18,22 with no strong correlation, except at 1000 Hz whose results were uncertain.

Conclusion:

We concluded that the average of the differences of hearing values obtained in Pure Tone Audiometry and Auditory Steady State Response were reasonably elevated mainly at 500 Hz, and at a lower level at 4000 Hz. Although the scores obtained in both tests had not shown strong correlation, they were slightly better at 500 Hz. The results for 1000 Hz are inconclusive regarding any existing correlation between these two tests.
RESUMEN

Introducción:

La audición, sentido que nos permite recibir la información sonora, se presenta como uno de los facilitadores para una interacción social eficaz. Un déficit en esta función acarrea perjuicio en el proceso comunicativo, lo que debe ser diagnosticado y tratado lo antes posible. Actualmente disponemos de medios subjetivos y objetivos para evaluar el sistema auditivo. La audiometría tonal es la forma subjetiva más aplicada para medir ese sentido. Sin embargo, en poblaciones imposibilitadas de responder subjetivamente, se aplican técnicas subjetivas como el Potencial Evocado Auditivo de Estado Estable. Ambas investigan las respuestas auditivas del sujeto bajo test, utilizando técnicas distintas.

Objetivo:

Evaluar a adultos jóvenes, sin quejas auditivas, investigando la sensibilidad auditiva y, a partir de los indicadores estadísticos, calcular los promedios de la diferencias de las respuestas obtenidas en esos procedimientos, y comprobar si existe correlación entre ellos.

Método:

Pesquisa de respuestas auditivas vía Audiometría Tonal y Potencial Evocado Auditivo de Estado Estable en las frecuencias de 500,1000, 2000 y 4000 Hz, en 30 sujetos oyentes normales.

Resultados:

Tras el cálculo del promedio de las diferencias entre las frecuencias testadas en los dos procedimientos, se pudo verificar una variación de 10,47 a 18,22, donde no se ha registrado fuerte correlación en ninguna de ellas, dejando duda en la frecuencia de 1000 Hz.

Conclusión:

Podemos decir que los promedios de las diferencias entre los valores obtenidos en los dos procedimientos fueron razonablemente elevados, sobre todo en 500 Hz y, en menor proporción, para 4kHz. Los valores que se obtuvieron en los dos exámenes, aunque no hayan presentado fuerte correlación, se presentaron discretamente mejores para 500 Hz. Resultados de 1000 Hz no nos permiten afirmar, incluso, si hay alguna correlación entre los testes.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Audiometry, Evoked Response / Audiometry, Pure-Tone / Auditory Threshold / Hearing / Interpersonal Relations Limits: Adolescent / Adult / Female / Humans / Male Language: Portuguese Journal: Distúrb. comun Journal subject: Speech-Language Pathology Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal do Rio de Janeiro/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Audiometry, Evoked Response / Audiometry, Pure-Tone / Auditory Threshold / Hearing / Interpersonal Relations Limits: Adolescent / Adult / Female / Humans / Male Language: Portuguese Journal: Distúrb. comun Journal subject: Speech-Language Pathology Year: 2018 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal do Rio de Janeiro/BR