Preliminary results of cervical impedance measurements
Res. Biomed. Eng. (Online)
;
34(2): 110-114, Apr.-June 2018. tab, graf
Article
in English
| LILACS
| ID: biblio-956294
ABSTRACT
AbstractIntroduction: In this work, we report the development and initial tests of a measuring system that is used to measure the cervical electrical impedance with the purpose of assessing cervical ripening. The objective is to offer a system that provides real-time information about the cervical electrical impedance and, indirectly, of the cervical ripening with the aim of preventing premature birth. Methods Probes were developed and initial tests were performed. The probes used had electrodes that are far apart, with the assumption that in such cases deeper layers of the cervical tissue are measured. Results Preliminary measurements were performed in a single session on a non-gravid uterus, using as excitation signal a sinusoidal wave with frequency varying from 3 kHz to 100 kHz. Results indicate, as expected, that the magnitude of the impedance is inversely proportional to the frequency, indicating capacitive impedance. Conclusion Preliminary results were very encouraging. However, further measurements must be performed to correlate the impedance to the cervical ripening.
Full text:
Available
Index:
LILACS (Americas)
Language:
English
Journal:
Res. Biomed. Eng. (Online)
Journal subject:
Engenharia Biomdica
Year:
2018
Type:
Article
Affiliation country:
Brazil
Institution/Affiliation country:
State University of Maringa/BR
Similar
MEDLINE
...
LILACS
LIS