contour of cementless femoral stem has minor effect on initial periprosthetic von Mises stress distribution: a 3- dimensonal finite element analysis
Saudi Medical Journal. 2009; 30 (7): 947-951
in English | IMEMR | ID: emr-103847
ABSTRACT
To evaluate periprosthetic von Mises stress distribution with cementless femoral stems of various contours. The study was carried out at the Department of Orthopaedics, Shanghai 6th Hospital, Shanghai, China between May 2008 and February 2009. Finite element models of proximal femoral replacement with 4 cementless stems [Alloclassic, Ribbed Anatomic, VerSys, and Securi-fit] of various contours were set up. Under the loading conditions of walking and stair climbing, 3-dimensional periprosthetic von Mises stresses were calculated, and the stress distribution patterns were compared. Periprosthetic stresses were increased in level 1, 2, and 3 under the 2 loading conditions, and more considerably in level 2 and 3. The stresses were higher on the medial side in all cases. No remarkable difference was found in the patterns between the 4 stems. The contour design of femoral stem has minor effect on initial periprosthetic von Mises stress distribution
Subject(s)
Search on Google
Index: IMEMR (Eastern Mediterranean) Main subject: Prostheses and Implants / Stress, Mechanical / Finite Element Analysis Limits: Humans Language: English Journal: Saudi Med. J. Year: 2009