Your browser doesn't support javascript.
loading
Does countersinking improve implant primary stability in low density bone?
Smile Dental Journal. 2012; 7 (1): 20-24
in English | IMEMR | ID: emr-144459
ABSTRACT
In order to evaluate the effect of countersinking on dental implants primary stability installed in low density bones, thirty dental implants were installed in the most proximal region of oxen ribs and grouped as follow Group1 Fifteen implants installed by standard drilling with countersinking, Group2 Fifteen implants installed by standard drilling without countersinking. The graduated manual torque wrench and wireless resonance frequency analyzer were used to record the peak insertion torque [IT] and implant stability quotient [ISQ] values respectively. The results revealed a significant difference in IT and ISQ values between the two groups of study with higher values recorded in group two. From these findings and with the limitation of this study, it was concluded that in low density bones, the avoidance of countersinking during implant bed preparation will significantly increases implant primary stability
Subject(s)
Search on Google
Index: IMEMR (Eastern Mediterranean) Main subject: Bone Density / Dental Implants Language: English Journal: Smile Dent. J. Year: 2012

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: IMEMR (Eastern Mediterranean) Main subject: Bone Density / Dental Implants Language: English Journal: Smile Dent. J. Year: 2012