[Standard full field electroretinography: normal values and variations with age in an Iranian population]
Bina Journal of Ophthalmology. 2008; 14 (1): 16-21
in Persian | IMEMR | ID: emr-165144
ABSTRACT
To determine normal values of standard full field electroretinography [ERG] and to evaluate variation with age in an Iranian population. The amplitudes and implicit times of ERGs from 170 normal subjects aged 1 to 80 years were recorded according to the International Society for Clinical Electrophysiology of Vision. Study participants included clients for spectacles, other patients' attendants and personnel of the hospital. Light-adapted ERG including single-white flash, 30-Hz flicker as well as dark-adapted ERG including rod, maximal darkadapted and cone responses were compiled. There were no significant differences between men and women, or between right and left eye in terms of amplitude or implicit times of ERG. An increase in amplitude and implicit time was observed in 70-80 year-old subjects as compared to younger individuals [P=0.04 and P=0.03, respectively]. ERG records are significantly diminished with advanced age. Our results are a suitable reference for evaluation of standard ERG responses
Search on Google
Index: IMEMR (Eastern Mediterranean) Language: Persian Journal: Bina J. Ophthalmol. Year: 2008