Your browser doesn't support javascript.
loading
Fetal heart rate monitoring by magnetocardiograms
Braz. j. med. biol. res ; 28(11/12): 1333-7, Nov.-Dec. 1995. graf
Article in English | LILACS | ID: lil-161535
ABSTRACT
The magnetic field produced by the fetal heart magnetocardiogram (FMCG) was measured by a single channel SQUID biogradiometer in an unshielded environment. FMCG amplitude ranged from 1 to 4 pT. Instantaneous fetal heart rate and averaged FMCG waveforms were computed. Spectral analysis was used to quantify fetal heart rate variability, and revealed peaks from 0.5 to 1.0 Hz, consistent with respiratory sinus arrhythmia. The duration of the QRS complex was approximately 50 msec, and its amplitude and shape showed strong spatial variation. Signal-to-noise ratio was inadequate for the study of P and T waves. These data show that this relatively simple and inexpensive instrument can be used for noninvasive in utero investigation of fetal heart activity without requiring a special environment.
Subject(s)
Search on Google
Index: LILACS (Americas) Main subject: Electrocardiography / Fetal Heart / Fetal Monitoring Limits: Female / Humans / Pregnancy Language: English Journal: Braz. j. med. biol. res Journal subject: Biology / Medicine Year: 1995 Type: Article / Congress and conference