ABSTRACT
A one-step enzyme linked sandwich immunoassay using Silicone rods coated with rabbit anti-human thyroglobulin anti-Tg) immunoglobulin G (Fab') conjugated with beta-D-galactosidase was established for the measurement of thyroglobulin in human sera. The volume of serum needed for the assay was 2 mul The sensitivity of the assay was 1.52 amol/tube, corresponding to O.5 ng/ml. The precision was proven by coefficients of variation: intra-assay, 7.0 to 9.1 per cent: inter-assay, 5.3 to 7.4 per cent. The correlation between this EIA and RIA was O.91, p < O.O1.