Análise química e estrutural de componente femoral de prótese total do quadril removido após fratura por fadiga: relato de caso / Chemical and structural analysis of total hip prosthesis femoral component retrieved after fatigue fracture: a case report
Rev. bras. ortop
;
37(6): 256-258, jun. 2002. ilus, tab
Article
in Portuguese
| LILACS
| ID: lil-322141
ABSTRACT
The authors present one case of stem fatigue fracture ofa cemented total hip prosthesis, which took place 43 months after its implantation. There was good cement thickness, adequate stem positioning, and absence of radiographic signs of loosening at the revision. The retrieved implant and an unused stem of the same type (control) were analyzed by macrographic inspection, metallographic and chemical analysis. The implants were found to have improper attributes of manufacturing and structural nature
Search on Google
Index:
LILACS (Americas)
Main subject:
Follow-Up Studies
/
Hip Fractures
Type of study:
Observational study
/
Prognostic study
Limits:
Aged
/
Humans
/
Male
Language:
Portuguese
Journal:
Rev. bras. ortop
Journal subject:
Orthopedics
Year:
2002
Type:
Article
Affiliation country:
Brazil
Institution/Affiliation country:
Instituto Nacional de Tecnologia/BR
/
Universidade Estaudl do Rio de Janeiro/BR