Comparação entre o polarímetro de varredura a laser, a tomografia de coerência óptica 1 e o Stratus-oct na detecção da perda axonal da atrofia em banda do nervo óptico / Comparison of scanning laser polarimetry, optical coherence tomography 1 and Stratus optical coherence tomography for the detection of axonal loss in band atrophy of the optic nerve
Leal, Bruno Campelo; Moura, Fredterico Castelo; Monteiro, Mário Luiz Ribeiro.
  • Leal, Bruno Campelo; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Divisão Oftalmológica. São Paulo. BR
  • Moura, Fredterico Castelo; Universidade de São Paulo. Faculdade de Medicina. Hospital das Clínicas. Divisão Oftalmológica. São Paulo. BR
  • Monteiro, Mário Luiz Ribeiro; USP. Faculdade de Medicina. Hospital das Clínicas. Divisão Oftalmológica. São Paulo. BR
Arq. bras. oftalmol ; 69(4): 531-537, jul.-ago. 2006. tab
Article in Portuguese, English | LILACS | ID: lil-435738
RESUMO

OBJETIVO:

Comparar a capacidade do polarímetro de varredura a laser (GDx), do tomógrafo de coerência óptica (OCT) 1 e do Stratus-OCT em diferenciar olhos normais de olhos com atrofia em banda do nervo óptico e hemianopsia temporal.

MÉTODOS:

Vinte e três olhos de pacientes com atrofia em banda do nervo óptico e 23 olhos de indivíduos normais foram incluídos em estudo prospectivo observacional caso-controle. Todos foram submetidos à análise da camada de fibras nervosas retiniana (CFNR) utilizando GDx, OCT-1 e Stratus-OCT. As médias dos valores obtidos em cada aparelho foram comparadas entre olhos com atrofia em banda e controles normais. Curvas ROC (receiver operating characteristic) e sensibilidade para especificidades fixas (80 por cento e 95 por cento) foram calculadas para cada parâmetro produzido pelos três instrumentos e comparadas entre si.

RESULTADOS:

Quando comparados aos indivíduos normais, os resultados dos pacientes referentes à camada de fibras nervosas retiniana (média global e quatro quadrantes estudados) foram significativamente menores (p<0,05) em todos os aparelhos estudados, com exceção do parâmetro referente ao setor temporal quando avaliado pelo GDx. A comparação das áreas sob a curva ROC (AROC) dos parâmetros referentes aos três aparelhos mostrou valores significativamente maiores para o Stratus-OCT quando comparados ao OCT-1 na média global e no setor temporal. O Stratus-OCT foi significativamente mais sensível do que o GDx na média global e nos quadrantes temporal, nasal e inferior. Por sua vez, o OCT-1 foi superior ao GDx na discriminação dos defeitos apenas no quadrante temporal, não mostrando diferença significativa na média global e nos demais quadrantes. Os três aparelhos mostraram capacidade semelhante na identificação dos defeitos referentes ao quadrante superior.

CONCLUSÃO:

O Stratus-OCT demonstrou a maior capacidade em diferenciar olhos com atrofia em banda do nervo óptico de olhos normais embora todos os três aparelhos consigam...
ABSTRACT

PURPOSE:

To compare the abilitiy of scanning laser polarimetry (GDx), optical coherence tomography (OCT) 1 and Stratus- optical coherence tomography to discriminate between healthy eyes and eyes with band atrophy of the optic nerve and temporal hemianopsia.

METHODS:

Twenty-three eyes with band atrophy of the optic nerve and 23 eyes from healthy subjects were included in this observational prospective case-control study. All eyes underwent retinal nerve fiber layer (RNFL) thickness analysis using GDx, optical coherence tomography-1 and Stratus-optical coherence tomography. Mean values obtained with each equipment were compared between band atrophy and normal eyes. Receiver operating characteristic (ROC) curves and sensitivities at fixed specificities (80 percent e 95 percent) were calculated for each parameter calculated with each equipment and compared.

RESULTS:

When compared with healthy subjects, the parameters corresponding to the global average as well as each of the four quadrant retinal nerve fiber layer thickness in eyes with band atrophy were significantly smaller (p<0.05), than in normal eyes, with the exception of GDx's temporal thickness parameter. Comparison of the areas under ROC curves (AUCs) of the parameters from the three equipments revealed significantly greater values for the Stratus-OCT when compared to the OCT-1 in the global average and in the temporal quadrant thickness measurement. Stratus-OCT was significantly more sensitive than GDx in the global average as well as in the temporal, nasal and inferior quadrant. OCT-1 was superior to GDx only in the temporal quadrant. All three equipments revealed a similar ability to identify retinal nerve fiber layer reduction in the superior quadrant.

CONCLUSIONS:

The Stratus OCT showed the best ability to discriminate between eyes with band atrophy of the optic nerve and healthy eyes although all three equipments were able do identify most of the abnormal eyes. OCT-1 was inferior to Stratus-OCT and...
Subject(s)
Full text: Available Index: LILACS (Americas) Main subject: Axons / Optic Atrophy / Tomography, Optical Coherence / Diagnostic Techniques, Ophthalmological Type of study: Diagnostic study / Observational study / Prognostic study Limits: Female / Humans / Male Language: English / Portuguese Journal: Arq. bras. oftalmol Journal subject: Ophthalmology Year: 2006 Type: Article Affiliation country: Brazil Institution/Affiliation country: USP/BR / Universidade de São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Axons / Optic Atrophy / Tomography, Optical Coherence / Diagnostic Techniques, Ophthalmological Type of study: Diagnostic study / Observational study / Prognostic study Limits: Female / Humans / Male Language: English / Portuguese Journal: Arq. bras. oftalmol Journal subject: Ophthalmology Year: 2006 Type: Article Affiliation country: Brazil Institution/Affiliation country: USP/BR / Universidade de São Paulo/BR