Avaliação tardia de endopróteses coronarianas com sirolimus: comparação da tomografia computadorizada por múltiplos detectores com a angiografia quantitativa e o ultra-som intracoronariano / Late outcome of sirolimus-eluting stents: comparison of multidetector computed tomography with quantitative coronary angiography and intravascular ultrasound
Pinto, Ibraim Masciarelli F; Sousa, Amanda G. M. R; Ishikama, Walther; Mattos, Luiz Alberto; Abizaid, Alexandre; Feres, Fausto; Tanajura, Luiz Fernando L; Sousa, Luiz C. B; Sousa, J. Eduardo M. R; Jatene, Adib.
  • Pinto, Ibraim Masciarelli F; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, Amanda G. M. R; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Ishikama, Walther; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Mattos, Luiz Alberto; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Abizaid, Alexandre; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Feres, Fausto; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Tanajura, Luiz Fernando L; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, Luiz C. B; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Sousa, J. Eduardo M. R; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
  • Jatene, Adib; Instituto Dante Pazzanese de Cardiologia. São Paulo. BR
Arq. bras. cardiol ; 87(5): 575-582, nov. 2006. ilus, graf
Article in Portuguese, English | LILACS, SES-SP | ID: lil-439701
RESUMO

OBJETIVO:

Avaliar os resultados da tomografia computadorizada por múltiplos detectores na avaliação dos resultados tardios de pacientes submetidos ao implante de endopróteses com sirolimus.

MÉTODOS:

Selecionamos 30 pacientes, previamente submetidos ao implante de stents com sirolimus com sucesso e com mais de seis meses de evolução. Todos foram submetidos à angiografia invasiva e ao ultra-som intravascular após a angiotomografia, feita com a injeção de 1,5 ml/kg de peso de meio de contraste iodado.

RESULTADOS:

A média dos diâmetros proximais de referência foi 3,01 ± 0,31 mm pela tomografia e 3,14 ± 0,31 mm pela angiografia (p = 0,04). Ao eliminarmos a artéria circunflexa da análise, a discrepância entre os dois exames deixou de ser significante -(tomografia= 3,01 ± 0,32 mm, angiografia= 3,10 ± 0,30 mm, p = 0,65). A média dos diâmetros distais de referência foi 2,86 ± 0,30 mm pela tomografia e 2,92 ± 0,32 pela angiografia (p = 0,25). A média do calibre mínimo no interior da endoprótese foi 2,85 ± 0,25 mm pela tomografia e 2,85 ± 0,29 mm angiografia (p = 0,27). A área de secção transversal mínima intra-stent foi 7,19 ± 1,47 mm² pela tomografia e 6,90 ± 1,52 mm² pelo ultra-som intracoronariano (p = 0,36), mas a correlação entre estas medidas era fraca (r= 0,33).

CONCLUSÃO:

A tomografia possibilita a avaliação qualitativa das endopróteses, a estimativa correta do diâmetro de referência proximal e distal dos vasos-alvo, além do calibre mínimo intra-stent. Sua correlação com as medidas feitas pelo ultra-som intracoronário, porém é menos intensa.
ABSTRACT

OBJECTIVE:

To assess the performance of multidetector computed tomography in determining late clinical outcomes of patients undergoing sirolimus-eluting stent implantation.

METHODS:

Thirty patients, successfully submitted to sirolimus-eluting stent implantation for more than six months, were selected to participate in the study. All underwent invasive angiography and intravascular ultrasound following CT angiography using iodinated contrast medium at a dose of 1.5 ml/kg.

RESULTS:

Mean proximal reference diameter was 3.01 ± 0.31 mm by tomography and 3.14 ± 0.31 mm by angiography (p = 0.04). When the left circumflex artery was excluded from the analysis, the difference between both examinations was no longer significant (tomography = 3.01 ± 0.32 mm; angiography = 3.10 ± 0.30 mm, p = 0.65). Mean distal reference diameter was 2.86 ± 0.30 mm by tomography and 2.92 ± 0.32 by angiography (p = 0.25). Mean in-stent minimal lumen diameter was 2.85 ± 0.25 mm by tomography and 2.85 ± 0.29 mm by angiography (p = 0.27). Mean minimal in-stent cross-sectional area was 7.19 ± 1.47 mm² by tomography and 6.90 ± 1.52 mm² by intravascular ultrasound (p = 0.36), but there was only a weak correlation between these measurements (r = 0.33).

CONCLUSION:

Computed tomography allows the qualitative assessment of sirolimus-eluting stents, accurate estimate of proximal and distal reference diameters of the target vessel, and in-stent minimal lumen diameter. Its correlation with measurements performed using intravascular ultrasound, however, is less strong.
Subject(s)
Full text: Available Index: LILACS (Americas) Main subject: Coronary Artery Disease / Angioplasty, Balloon, Coronary / Stents / Sirolimus Type of study: Observational study / Prognostic study / Qualitative research Limits: Female / Humans / Male Language: English / Portuguese Journal: Arq. bras. cardiol Year: 2006 Type: Article Institution/Affiliation country: Instituto Dante Pazzanese de Cardiologia/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Coronary Artery Disease / Angioplasty, Balloon, Coronary / Stents / Sirolimus Type of study: Observational study / Prognostic study / Qualitative research Limits: Female / Humans / Male Language: English / Portuguese Journal: Arq. bras. cardiol Year: 2006 Type: Article Institution/Affiliation country: Instituto Dante Pazzanese de Cardiologia/BR