RESUMO

INTRODUÇÃO:

A avaliação de perfis hormonais exige, tradicionalmente, a participação de especialistas para adequada liberação setorial. A análise desses resultados através de regras booleanas é alternativa interessante para minimizar os resultados a serem avaliados manualmente. MATERIAL E

MÉTODO:

Avaliamos a utilização de regras booleanas construídas no programa Instrument Manager para o equipamento Architect, principalmente para mensuração de hormônios sexuais e tireoidianos, além de marcadores tumorais. O resultado da intervenção foi avaliado quanto a a) número e facilidade de construção das regras; b) comparação cega entre liberação por patologista clínico (resultados impressos) e conjunto de regras em 940 testes consecutivos.

RESULTADOS:

A criação das regras se revelou tarefa complexa e trabalhosa, especialmente pela existência de perfis hormonais com diversos modos de solicitação dos testes. Foram necessárias 153 regras booleanas (do tipo se...então) em uma ordem específica. Essas regras concordaram com o especialista em 97,9 por cento dos casos (920 testes). O conjunto de regras reteve 25 casos (2,7 por cento) e o patologista clínico, apenas nove. Nos 20 casos discordantes, as regras deixaram de reter apenas dois casos, sendo um da fração beta da gonadotrofina coriônica humana (beta-hCG) em homens (que implicou a criação de nova regra) e um caso de perfil tireoidiano completo sem hormônio tireoestimulante (TSH) (o patologista clínico optou por confirmar o pedido médico).

CONCLUSÃO:

A criação de conjunto eficiente de regras booleanas é tarefa complexa que necessita de conhecimentos técnicos e de lógica, mas que permite a otimização do funcionamento do laboratório. Obtivemos excelente concordância entre o conjunto de regras criado e a liberação manual de patologista clínico, garantindo a segurança, a velocidade e o menor custo do sistema.
ABSTRACT

BACKGROUND:

Hormone profile evaluation traditionally requires an expert review of results for adequate laboratory section test release. Analysis of these results using Boolean rules is an interesting alternative to reduce the number of results that require manual review. MATERIAL AND

METHOD:

We evaluated the utilization of Boolean rules using Instrument Manager software and Architect analyzer, mainly performing sex and thyroid hormones measurement. The intervention was evaluated on a) number of rules and its easiness of construction; b) blind comparison of results evaluation by clinical pathologist (printed results) and set of rules in 940 consecutive tests.

RESULTS:

Rule creation was a complex and arduous task, especially due to hormonal profiles with several different request patterns. It was necessary to use a set of 153 Boolean (if...then) rules, in a specific order. This set of rules agreed with expert opinion in 97.9 percent (920 tests). Rules hold 25 tests (2.7 percent) and the clinical pathologist only nine tests. There was discordance in 20 cases; rules did not hold only two cases a beta-hCG in a male patient (that prompted the creation of a new rule) and a complete thyroid profile lacking only TSH request (pathologist opted to review the original request).

CONCLUSION:

Creation of an efficient set of Boolean rules proved to be a complex task requiring both technical and logics knowledge, but allowing optimization of laboratory workload. We achieved excellent concordance between the set of rules and clinical pathologist manual review, in a safe, fast and low cost system.
Full text: Available Index: LILACS (Americas) Type of study: Practice guideline Language: Portuguese Journal: J. bras. patol. med. lab Journal subject: Pathology Year: 2006 Type: Article Affiliation country: Brazil Institution/Affiliation country: SCSP/BR / Santa Casa de São Paulo/BR / Universidade São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Type of study: Practice guideline Language: Portuguese Journal: J. bras. patol. med. lab Journal subject: Pathology Year: 2006 Type: Article Affiliation country: Brazil Institution/Affiliation country: SCSP/BR / Santa Casa de São Paulo/BR / Universidade São Paulo/BR