Bomba sangüínea espiral: concepção, desenvolvimento e aplicação clínica de projeto original / Spiral blood pump: conception, development and clinical application of the original project
Dinkhuysen, Jarbas J; Andrade, Aron José Pazin de; Manrique, Ricardo; Saito, Claudia Sanches Medina; Leme, Juliana; Biscegli, Francisco.
  • Dinkhuysen, Jarbas J; Instituto Dante Pazzanese de Cardiologia. Seção Médica de Transplante. São Paulo. BR
  • Andrade, Aron José Pazin de; IDPC/SP. Divisão de Bioengenharia. São Paulo. BR
  • Manrique, Ricardo; IDPC/SP. Laboratório Experimental. São Paulo. BR
  • Saito, Claudia Sanches Medina; IDPC/SP. Departamento Cirurgia Cardiovascular. São Paulo. BR
  • Leme, Juliana; IDPC/SP. Divisão de Bioengenharia. São Paulo. BR
  • Biscegli, Francisco; IDPC/SP. Divisão de Bioengenharia. São Paulo. BR
Rev. bras. cir. cardiovasc ; 22(2): 224-234, abr.-jun. 2007. ilus, tab, graf
Article in Portuguese | LILACS, SES-SP | ID: lil-461764
RESUMO

INTRODUÇÃO:

O trabalho aborda projeto original relativo à concepção, ao desenvolvimento e à aplicação clínica de bomba sangüínea que associa forças centrífuga e axial de propulsão hidráulica, baseada no princípio de Arquimedes, denominada Bomba Espiral (BE), tendo recebido Patente Nacional e Relatório Preliminar Internacional categorizando-a como invento.

MÉTODO:

Visa avaliar sua capacidade hidrodinâmica e seu impacto aos elementos figurados do sangue por meio de testes "in vitro", como Eficiência Hidrodinâmica, Hemólise Normalizada e Visibilização de Escoamento, e, nos testes "in vivo" experimentais, feitos em carneiros submetidos a Circulação Extracorpórea (CEC), e clínico, em 43 pacientes submetidos a operações cardíacas com CEC, nas quais o elemento propulsor foi a BE.

RESULTADOS:

Na dependência da distância entre o rotor e a carcaça (fenda) da bomba pôde-se observar que com 1,5 mm gerou escoamento ao redor de 9 L/min, pressão acima de 400mmHg com 1500 rotações por minuto (rpm), índices de Hemólise Normalizada não superiores a 0,0375 g/100l em condições de alto fluxo e pressão, e pelo estudo de Visibilização do Escoamento no interior da bomba não se detectou áreas de estagnação ou turbulência na entrada, saída e junto à extremidade dos fusos. Nas pesquisas "in vivo" experimentais em ovinos em CEC por 6 horas a BE foi capaz de manter parâmetros pressóricos adequados e Hemoglobina Livre entre 16,36 mg por cento e 44,90 mg por cento. Durante sua aplicação em cirurgias cardíacas com CEC, num grupo de 43 pacientes, pôde-se constatar variações pré e pós-CEC, na Hemoglobina Livre de 9,34 a 44,16 mg por cento, no Fibrinogênio, de 236,65 a 547,26 mg por cento, na contagem do número de Plaquetas de 152,465 a 98,139, Desidrogenase Láctica, de 238,12 a 547,26 mg por cento, com tempo de coagulação ativada ao redor de 800 seg. quando em CEC.

CONCLUSÕES:

A BE mostrou resolutividade por gerar escoamento e pressão adequados, sem causar danos excessivos...
ABSTRACT

INTRODUCTION:

This paper addresses an original project that encompasses the conception, development and clinical application of a helical bypass pump called the Spiral Pump, that uses the association of centrifugal and axial propulsion forces based de the Archimedes principle. This project has obtained a Brazilian Patent and an International Preliminary Report, defining it as an invention.

METHODS:

The aim of this work was to evaluate the hemodynamic capacity and the impact of its application on blood cells by means of experimental in vitro tests, including hydrodynamic efficiency, effect on hemolysis and flow visualization. Moreover, in vivo experimental tests were carried out on lambs that were submitted to cardiopulmonary bypass for six hours and in 43 patients submitted to heart bypass surgery using the Spiral Pump.

RESULTS:

When the rotor-plastic casing gap was 1.5mm, the flow generated was nearly 9 L/min, the pressure was greater than 400 mmHg at 1500 rpm, and the normalized hemolytic indexes were not greater than 0.0375 g/100L in high-flow and pressure conditions. Additionally, by the flow visualization techniques, stagnation was not seen inside the pump nor was turbulence identified at the entrance or exit of the pump, or at the ends of the spindles. In the in vivo tests using cardiopulmonary bypasses for 6 hours in lambs, the pump maintained adequate pressure rates and the free hemoglobin levels ranged between 16.36 mg percent and 44.90 mg percent. Evaluating the results of the 43 patients who used this pump in heart bypass operations we observed that the free hemoglobin ranged from 9.34 mg percent before to 44.16 mg percent after surgery, the serum fibrinogen was from 236.65 mg percent to 547.26mg percent, platelet blood count from 152,465 to 98,139 and the lactic dehydrogenase from 238.12mg percent to 547.26mg percent. The Activated Coagulation Time was close to 800 seconds during the bypass.

CONCLUSION:

The Spiral Pump was...
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: In Vitro Techniques / Cardiovascular Diseases / Heart-Assist Devices / Hemolysis Limits: Animals / Humans Language: Portuguese Journal: Rev. bras. cir. cardiovasc Year: 2007 Type: Article Institution/Affiliation country: SP+BR / Instituto Dante Pazzanese de Cardiologia/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: In Vitro Techniques / Cardiovascular Diseases / Heart-Assist Devices / Hemolysis Limits: Animals / Humans Language: Portuguese Journal: Rev. bras. cir. cardiovasc Year: 2007 Type: Article Institution/Affiliation country: SP+BR / Instituto Dante Pazzanese de Cardiologia/BR