RESUMO
OBJETIVO:
Avaliar a prótese Polyflex® quanto à sua eficácia, facilidade de implantação e complicações em pacientes com afecções traqueobrônquicas.MÉTODOS:
Foram acompanhados, prospectivamente, dezesseis pacientes com estenoses traqueais secundárias à intubação orotraqueal (n = 12), neoplasia (n = 3) e granulomatose de Wegener (n = 1), não candidatos a tratamento cirúrgico. Desses, onze eram mulheres e cinco eram homens, com idade média de 42,8 anos (intervalo de 21 a 72 anos). Os pacientes foram submetidos à implantação de um total de 21 próteses Polyflex®. Os procedimentos foram realizados no centro cirúrgico, sob anestesia geral e as próteses implantadas através de laringoscopia de suspensão e aplicador próprio.RESULTADOS:
Em todos os casos foi possível implantar a prótese e observamos resolução de sintomas. Os pacientes permaneceram com a prótese por tempo médio de 7,45 meses, variando entre 2 e 18 meses. As complicações pós-operatórias imediatas observadas foram disfonia em dois pacientes (12,5 por cento) e odinofagia em dois pacientes (12,5 por cento). As complicações tardias foram tosse em dez pacientes (62,5 por cento), migração em sete pacientes (43,75 por cento), formação de granulomas em dois pacientes (12,5 por cento) e pneumonia em um paciente (6,25 por cento).CONCLUSÃO:
A prótese Polyflex® é fácil de implantar e retirar, é bem tolerada e efetiva na resolução dos sintomas, porém, está associada a alto índice de migração, principalmente em estenoses pós-intubação orotraqueal.ABSTRACT
OBJECTIVE:
To evaluate the Polyflex® stent in terms of its efficacy, ease of implantation, and complications in patients with tracheobronchial affections.METHODS:
This was a prospective study, in which sixteen patients with inoperable tracheal stenosis secondary to orotracheal intubation (n = 12), neoplasia (n = 3), or Wegener's granulomatosis (n = 1) were monitored. Of these patients, eleven were women, and five were men. The mean age was 42.8 years (range, 21-72 years). Patients were submitted to implantation of a total of 21 Polyflex® stents. All procedures were carried out in the operating room under general anesthesia, and the stents were implanted via suspension laryngoscopy using the stent applicator.RESULTS:
Stents were implanted and symptoms were resolved in all cases. The stents remained in place for a mean period of 7.45 months, ranging from 2 to 18 months. The complications observed in the immediate postoperative period were dysphonia (in two patients, 12.5 percent) and odynophagia (in two patients, 12.5 percent). Late complications were cough (in ten patients, 62.5 percent), migration (in seven patients, 43.75 percent), granuloma formation (in two patients, 12.5 percent), and pneumonia (in one patient, 6.25 percent).CONCLUSION:
The Polyflex® stent is easily implanted, easily removed, well tolerated by patients and effective in resolving symptoms. However, its use is associated with a high rate of migration, especially in patients with post-orotracheal intubation stenosis.
Full text:
Available
Index:
LILACS (Americas)
Main subject:
Silicones
/
Tracheal Stenosis
/
Stents
Type of study:
Etiology study
/
Evaluation studies
/
Observational study
/
Prognostic study
/
Risk factors
Limits:
Adult
/
Aged
/
Female
/
Humans
/
Male
Language:
Portuguese
Journal:
J. bras. pneumol
Journal subject:
Pulmonary Disease (Specialty)
Year:
2007
Type:
Article
Affiliation country:
Brazil
Institution/Affiliation country:
Universidade de São Paulo/BR
Similar
MEDLINE
...
LILACS
LIS