Alterações da pressão de pico inspiratório e do volume corrente fornecidos por reanimadores manuais com balão auto-inflável em função do fluxo de entrada de oxigênio utilizado / Alterations in peak inspiratory pressure and tidal volume delivered by manually operated self-inflating resuscitation bags as a function of the oxygen supply rate
Godoy, Armando Carlos Franco de; Vieira, Ronan José; De Capitani, Eduardo Mello.
  • Godoy, Armando Carlos Franco de; Universidade Estadual de Campinas. Hospital das Clínicas. Serviço de Emergência. Campinas. BR
  • Vieira, Ronan José; Universidade Estadual de Campinas. Hospital das Clínicas. Serviço de Emergência. Campinas. BR
  • De Capitani, Eduardo Mello; Universidade Estadual de Campinas. Hospital das Clínicas. Departamento de Doenças Pulmonares. Campinas. BR
J. bras. pneumol ; 34(10): 817-821, out. 2008. ilus, tab
Article in English, Portuguese | LILACS | ID: lil-496617
RESUMO

OBJETIVO:

Determinar possíveis alterações do volume corrente e da pressão de pico inspiratório fornecidos por sete modelos de reanimador manual com balão auto-inflável em função do fluxo de entrada de oxigênio utilizado.

MÉTODOS:

Os reanimadores testados foram Oxigel, modelos A e B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; e Narcosul. Para as aferições, acoplaram-se um fluxômetro de oxigênio de parede, um fluxômetro/respirômetro, um reanimador, um sensor (aparelho Tracer 5) e um pulmão-teste. Além disso, acoplou-se o aparelho Tracer 5 a um notebook. Utilizaram-se fluxos de entrada de oxigênio de 1, 5, 10 e 15 L/min.

RESULTADOS:

O volume corrente fornecido pelo reanimador Oxigel modelo A ao receber 15 L/min de oxigênio foi aproximadamente 99 por cento maior que o fornecido ao receber 1 L/min de oxigênio. Da mesma forma, a pressão de pico inspiratório foi 155 por cento maior. Nas mesmas condições, o volume corrente fornecido pelo reanimador Narcosul foi 48 por cento maior, e a pressão de pico inspiratório foi 105 por cento maior. Os demais reanimadores testados não apresentaram alterações significativas do volume corrente e da pressão de pico inspiratório fornecidos.

CONCLUSÕES:

Nas condições de resistência e complacência utilizadas, os reanimadores em que o fluxo de entrada de oxigênio é direcionado diretamente ao interior do balão tiveram a válvula do paciente presa em posição de inspiração ao receberem um fluxo > 5 L/min, aumentando significativamente o volume corrente e a pressão de pico inspiratório fornecidos. Isso não ocorreu nos reanimadores em que o fluxo de entrada de oxigênio é direcionado ao exterior do balão.
ABSTRACT

OBJECTIVE:

To assess possible alterations in the tidal volume and peak inspiratory pressure delivered by seven models of manually operated self-inflating resuscitation bags as a function of the oxygen supply rate used.

METHODS:

The resuscitation bags tested were the following Oxigel, models A and B; CE Reanimadores; ProtecSolutions; Missouri; Axmed; and Narcosul. For the measurements, a wall oxygen flow meter, a flow meter/respirometer, a resuscitation bag, a sensor (Tracer 5 unit), and a test lung were connected. In addition, the Tracer 5 unit was connected to a notebook computer. Oxygen supply rates of 1, 5, 10, and 15 L/min were used.

RESULTS:

The tidal volume delivered by the Oxigel model A resuscitation bag when receiving oxygen at a rate of 15 L/min was approximately 99 percent greater than that delivered when receiving oxygen at a rate of 1 L/min. Similarly, peak inspiratory pressure was approximately 155 percent greater. Under the same conditions, the tidal volume delivered by the Narcosul resuscitation bag was 48 percent greater, and peak inspiratory pressure was 105 percent greater. The remaining resuscitation bags tested showed no significant alterations in the tidal volume or peak inspiratory pressure delivered.

CONCLUSIONS:

Under the resistance and compliance conditions used, the resuscitation bags in which the oxygen inflow is directly to the interior of the bag had the patient valve stuck at the inspiratory position when receiving oxygen at a rate > 5 L/min, significantly increasing the tidal volume and peak inspiratory pressure delivered. This did not occur with the resuscitation bags in which the oxygen inflow is directed to the exterior of the bag.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Oxygen / Respiration, Artificial / Resuscitation / Ventilators, Mechanical / Tidal Volume / Inhalation Type of study: Evaluation studies / Prognostic study Limits: Humans Language: English / Portuguese Journal: J. bras. pneumol Journal subject: Pulmonary Disease (Specialty) Year: 2008 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Estadual de Campinas/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Oxygen / Respiration, Artificial / Resuscitation / Ventilators, Mechanical / Tidal Volume / Inhalation Type of study: Evaluation studies / Prognostic study Limits: Humans Language: English / Portuguese Journal: J. bras. pneumol Journal subject: Pulmonary Disease (Specialty) Year: 2008 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Estadual de Campinas/BR