Article in English, Portuguese | LILACS | ID: lil-514838
RESUMO
Em crianças usuárias de Implante Coclear (IC), o registro do potencial de ação composto do nervo auditivo evocado eletricamente (ECAP) representa uma maneira de avaliar as modificações nas respostas neurais e entre o feixe de eletrodos e o tecido neural ao longo do tempo.

OBJETIVO:

Estudar o ECAP em crianças ao longo do primeiro ano de uso do IC. MATERIAL E

MÉTODO:

As características do ECAP foram analisadas em 13 crianças, implantadas com idades inferiores a três anos de idade. Estudo de série.

RESULTADOS:

Houve aumento estatisticamente significante na amplitude do pico N1, nos eletrodos basais, entre o segundo e o terceiro retorno. Não foram obtidas diferenças significantes para a latência do pico N1, o slope, o p-NRT e o período de recuperação entre os retornos. Nos três retornos, a maioria dos sujeitos apresentou tempo de recuperação entre 1000 e 2000µs.

CONCLUSÃO:

Ao longo do primeiro ano de uso do IC, a estimulação elétrica liberada pelos eletrodos intracocleares não causou alterações significativas às características do ECAP, exceto pelo aumento da amplitude do pico N1.
ABSTRACT
In children with cochlear implant (CI), the recording of the electrically evoked compound action potential (ECAP) of the auditory nerve represents an option to assess changes in auditory nerve responses and the interaction between the electrode bundle and the neural tissue over time.

AIM:

To study ECAP in children during the first year of CI use. MATERIALS AND

METHODS:

The ECAP characteristics have been analyzed in 13 children implanted younger than three years of age. Series study.

RESULTS:

During the first year of CI use there was a significant statistical raise in the N1 peak amplitude, in basal electrodes, between the second and third return visits. There were not any significant differences obtained for N1 peak, latency, slope, p-NRT or recovery time, in the return visits.

CONCLUSION:

During the first year of CI use, the electrical stimulation provided by the intracochlear electrodes did not cause significant changes to ECAP characteristics, except for an increase in N1 peak amplitude.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Evoked Potentials, Auditory, Brain Stem / Cochlear Implants / Cochlear Nerve / Hearing Loss, Sensorineural Type of study: Observational study Limits: Child, preschool / Humans / Infant Language: English / Portuguese Journal: Rev. bras. otorrinolaringol Journal subject: Otolaryngology Year: 2009 Type: Article Affiliation country: Brazil

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Evoked Potentials, Auditory, Brain Stem / Cochlear Implants / Cochlear Nerve / Hearing Loss, Sensorineural Type of study: Observational study Limits: Child, preschool / Humans / Infant Language: English / Portuguese Journal: Rev. bras. otorrinolaringol Journal subject: Otolaryngology Year: 2009 Type: Article Affiliation country: Brazil