Determinação das glicemias capilar e venosa com glicosímetro versus dosagem laboratorial da glicose plasmática / Determination of capillary blood glucose and venous blood glucose with a glucometer versus laboratory determination of venous plasma glucose
Cordova, Caio Mauricio Mendes de; Valle, Jéssyca Pereira; Yamanaka, Celina Noriko; Cordova, Maurício Mendes de.
  • Cordova, Caio Mauricio Mendes de; Universidade Regional de Blumenau. Departamento de Ciências Farmacêuticas. BR
  • Valle, Jéssyca Pereira; s.af
  • Yamanaka, Celina Noriko; s.af
  • Cordova, Maurício Mendes de; s.af
J. bras. patol. med. lab ; 45(5): 378-384, out. 2009. ilus, tab
Article in Portuguese | LILACS | ID: lil-536888
RESUMO
INTRODUÇÃO E OBJETIVOS: Diabetes mellitus (DM) é a mais importante patologia que envolve o pâncreas endócrino, sendo uma das principais causas de morbidade e mortalidade na população geral. O objetivo deste trabalho foi avaliar a determinação da glicemia em diferentes tipos de amostras e metodologias. MÉTODOS: Utilizando um equipamento Accu-Check Advantage (Roche), foi avaliada a glicemia em amostras de sangue capilar (GCG) e de sangue venoso (GVG), e a dosagem da glicemia venosa em plasma (GVE) foi realizada por método enzimático de rotina. RESULTADOS E CONCLUSÃO: Foi observada boa correlação entre a GCG e a GVG (r = 0,8742). Entretanto, houve diferença significativa entre a GCG e a GVE (r = 0,6543) e entre a GVG e a GVE (r = 0,5038) (p < 0,001). Essa diferença é maior considerando-se apenas os pacientes normoglicêmicos. O fato de haver diferentes estudos com resultados conflitantes entre si não depende somente da marca ou série específica do glicosímetro, já que uma mesma marca pode apresentar resultados inconsistentes em diferentes estudos.
ABSTRACT
INTRODUCTION AND OBJECTIVES: Diabetes mellitus is the most important pathology that affects the endocrine pancreas and one of the main causes of morbidity and mortality among the general population. The aim of this work was to evaluate the correlation of blood glucose determination in different types of samples and methodologies. METHODS: Using an Accu-Check Advantage glucometer (Roche), it was determined the blood glucose levels in capillary (CBG) and venous (VBG) samples, and plasma venous glucose was determined by routine enzymatic methodology (EVG). RESULTS AND CONCLUSION: It was observed a good correlation between CBG and VBG (r = 0.8742). However, there was a significant difference between CBG and EVG (r = 0.6543) and between VBG and EVG (r = 0.5038) (p < 0.001). These differences are even more substantial considering only normoglycemic patients. The existence of different studies with inconsistent results does not depend only on the glucometer brand or its specific series, as the same device may present inconsistent results in different studies.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Blood Glucose / Blood Glucose Self-Monitoring / Diabetes Mellitus Type of study: Diagnostic study Limits: Adult / Female / Humans / Male Language: Portuguese Journal: J. bras. patol. med. lab Journal subject: Pathology Year: 2009 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Regional de Blumenau/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Blood Glucose / Blood Glucose Self-Monitoring / Diabetes Mellitus Type of study: Diagnostic study Limits: Adult / Female / Humans / Male Language: Portuguese Journal: J. bras. patol. med. lab Journal subject: Pathology Year: 2009 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Regional de Blumenau/BR