Your browser doesn't support javascript.
loading
Equipamento brasileiro de detecção gama intra-operatória para biópsia de linfonodo sentinela / Brazilian gamma detection device for sentinel lymph node biopsy
Oliveira Filho, Renato Santos de; Farah, Andréia Bufoni; Costa, Fábio Eduardo da; Nahas, Fabio Xerfan; Oliveira, Daniel Arcuschin de; Oliveira, Renan Gianotto de.
Affiliation
  • Oliveira Filho, Renato Santos de; Hospital Israelita Albert Einstein. São Paulo. BR
  • Farah, Andréia Bufoni; Hospital Israelita Albert Einstein. São Paulo. BR
  • Costa, Fábio Eduardo da; Instituto de Pesquisas Energéticas e Nucleares. São Paulo. BR
  • Nahas, Fabio Xerfan; Universidade Federal de São Paulo. São Paulo. BR
  • Oliveira, Daniel Arcuschin de; Universidade Anhembi Morumbi. Escola de Medicina. São Paulo. BR
  • Oliveira, Renan Gianotto de; Universidade Anhembi Morumbi. Escola de Medicina. São Paulo. BR
Rev. Col. Bras. Cir ; 37(1): 027-030, ene.-feb. 2010. ilus, tab
Article in Pt | LILACS | ID: lil-554488
Responsible library: BR1.1
RESUMO

OBJETIVO:

Testar a eficácia de equipamento de detecção gama intra-operatória (DGI) desenvolvido pelo IPEN (Brasil), em procedimentos de biópsia de linfonodo sentinela (BLS) no melanoma e no câncer de mama.

MÉTODOS:

Foram estudados 40 pacientes portadores de melanoma ou câncer de mama com indicação para realização de BLS.Todos pacientes foram submetidos à linfocintilografia e a BLS ocorreu entre 2 a 24 horas após a mesma. Concomitantemente à DGI, realizou-se o mapeamento linfático com corante vital. Foram feitas leituras com o equipamento convencional Neoprobe® 1500 e com o equipamento em teste (IPEN) dos valores de captação do sítio de injeção do radiofármaco, do LS in vivo e ex vivo e da captação de fundo. Foi registrado se o LS estava corado e se o cirurgião teve facilidade para encontrá-lo. Nos primeiros 20 pacientes utilizou-se o equipamento convencional e depois o de teste; nos outros 20, utilizou-se primeiro o equipamento em teste, com objetivo de verificar se o mesmo identificava primariamente o LS.

RESULTADOS:

Dos quarenta pacientes, 33 eram portadores de tumor de mama e sete de melanoma cutâneo; variação da idade 21 a 68 anos (mediana= 46 anos); 35 mulheres e 5 homens. Em apenas um paciente o LS não foi encontrado, nem pela DGI nem pelo corante vital. Não houve diferença estatística entre as razões ex vivo/fundo obtidas com os dois equipamentos (p=0, 2583-ns).

CONCLUSÃO:

É possível realizar o procedimento de BLS com o equipamento brasileiro desenvolvido pelo IPEN, com facilidade e sem prejuízo para o paciente.
ABSTRACT

OBJECTIVE:

To test the effectiveness of an intra-operative gamma detection Brazilian device (IPEN) on sentinel lymph node biopsy (SLNB) procedures.

METHODS:

Forty melanoma or breast cancer patients with indication for undergoing SLNB were studied. Lymphoscintigraphy was done 2 to 24 hours prior to surgery. Lymphatic mapping with vital dye and gamma detection were performed intraoperatively. For gamma detection Neoprobe ® 1500 was used followed by IPEN (equipment under test) in the first 20 patients and for the remaining half IPEN was used first to verify its ability to locate the sentinel node (SN). Measurements were taken from the radiopharmaceutical product injection site, from SN (in vivo and ex vivo) and from background. It was recorded if the SN was stained or not and if it was found easily by surgeon.

RESULTS:

There were 33 (82.5 percent) breast cancer and 7 (17.5 percent) melanoma patients. Ages varied from 21 to 68 year-old (median age of 46). Sex distribution was 35 (87.5 percent) women and 5 (12.5 percent) men. Sentinel node was found in all but one patient. There was no statistical difference between the reasons ex vivo/ background obtained with the measures of both equipments (p=0, 2583-ns). The SN was easily found by the surgeon with both devices.

CONCLUSION:

The SLNB was successfully performed using either equipment. It was possible to do SLNB with the Brazilian device developed by IPEN without prejudice for the patient.
Subject(s)
Key words
Full text: 1 Index: LILACS Main subject: Skin Neoplasms / Breast Neoplasms / Sentinel Lymph Node Biopsy / Gamma Rays / Intraoperative Care / Melanoma Type of study: Diagnostic_studies / Observational_studies / Prognostic_studies Limits: Adult / Aged / Humans Country/Region as subject: America do sul / Brasil Language: Pt Journal: Rev. Col. Bras. Cir Journal subject: CIRURGIA GERAL Year: 2010 Type: Article
Full text: 1 Index: LILACS Main subject: Skin Neoplasms / Breast Neoplasms / Sentinel Lymph Node Biopsy / Gamma Rays / Intraoperative Care / Melanoma Type of study: Diagnostic_studies / Observational_studies / Prognostic_studies Limits: Adult / Aged / Humans Country/Region as subject: America do sul / Brasil Language: Pt Journal: Rev. Col. Bras. Cir Journal subject: CIRURGIA GERAL Year: 2010 Type: Article