Your browser doesn't support javascript.
loading
Estudo normativo do eletrorretinograma por padrões (PERG) em adultos / Pattern-reversal electroretinogram (PERG): a normative study in adults
Alves, Letícia Dourado; Berezovsky, Adriana; Sacai, Paula Yuri; Pereira, Josenilson Martins; Salomão, Solange Rios.
  • Alves, Letícia Dourado; Universidade Federal de São Paulo. Departamento de Oftalmologia. São Paulo. BR
  • Berezovsky, Adriana; Universidade Federal de São Paulo. Departamento de Oftalmologia. São Paulo. BR
  • Sacai, Paula Yuri; Universidade Federal de São Paulo. São Paulo. BR
  • Pereira, Josenilson Martins; s.af
  • Salomão, Solange Rios; Universidade Federal de São Paulo. Departamento de Oftalmologia. São Paulo. BR
Arq. bras. oftalmol ; 73(3): 225-230, jun. 2010. ilus, graf, tab
Article in Portuguese | LILACS | ID: lil-555061
RESUMO

Objetivo:

Determinar valores para o eletrorretinograma por padrões (PERG) em voluntários adultos saudáveis seguindo o protocolo padrão recomendado pela International Society for Clinical Electrophysiology of Vision-ISCEV.

Métodos:

Os participantes foram 30 voluntários saudáveis (15 homens), com idade variando de 18 a 61 anos (média= 30,8 ± 8,7 anos, mediana= 29,6 anos). Os critérios de inclusão foram acuidade visual 0,0 logMAR (20/20 Snellen) em cada olho separadamente, ausência de queixas visuais, ausência de opacidades de meio, ausência de história pregressa de doenças oculares ou neurológicas, história familiar negativa para doenças oculares e assinatura de termo de consentimento livre e esclarecido para participação em pesquisa. O PERG foi registrado em sala escura a 1 metro de distância de um monitor de alta resolução. Os estímulos foram padrões monocromáticos em forma de tabuleiro de xadrez subtendendo 60, 15 e 7,5 minutos de arco revertendo a 1,9 Hz. As respostas foram obtidas com variante de eletrodo descartável de filamento desenvolvido pela UNIFESP. Latência (milissegundos) para os componentes N35, P50 e N95 bem como a amplitude (microvolts) pico a pico para o N35-P50 e P50-N95 foram determinadas. Para estudo normativo apenas um olho escolhido ao acaso foi incluído. Os limites normais foram calculados pelo percentil 97,5 por cento para latência e 2,5 por cento para amplitude em cada tamanho de estímulo.

Resultados:

Valores normativos para latência de N35, P50 e N95 para os estímulos de 60', 15' e 7,5' foram respectivamente N35 - 40,1; 39,9 e 41,3 ms; P50 - 60,5; 64,4 e 65,6 ms; N95 - 103,4; 104,6 e 104,6 ms. Para amplitude os valores normativos para os estímulos 60', 15' e 7,5' foram respectivamente N35-P50 - 1,7; 1,6 e 0,9 µV; P50-N95 - 3,8; 2,8 e 1,5 µV. Não houve diferenças entre os sexos para os três estímulos empregados tanto para latência como para amplitude. Não encontramos correlação entre os parâmetros estudados...
ABSTRACT

Purpose:

To determine normative values for pattern-reversal electroretinogram (PERG) in healthy adult volunteers according to the standard protocol recommended by the International Society for Clinical Electrophysiology of Vision-ISCEV.

Methods:

Participants were 30 healthy volunteers (15 males) with ages from 18 to 61 years (mean= 30.8 ± 8.7 yrs.; median= 29.5 yrs). Inclusion criteria were visual acuity of 0.0 logMAR (20/20 Snellen) in each eye, absence of visual complaints, absence of media opacities, negative history for ocular or neurological diseases, negative family history for ocular and informed consent. PERG was recorded from each eye in a darkened room at 1 m from a high resolution display monitor. Stimuli were monochromatic checkerboards subtending 60, 15 e 7.5 minutes of visual angle reversing at 1.9 Hz. Responses were obtained from modified disposable fiber electrodes developed at UNIFESP. Latency (ms) for N35, P50 and N95 components as well as peak-to-peak amplitudes (µV) for N35-P50 and P50-N95 were determined. For normative values only one randomly chosen eye was included. Normal limits were calculated as 97.5 percent percentiles for latency and 2.5 percent percentile for amplitudes for each stimulus size.

Results:

Normal limits for N35, P50 and N95 latencies for 60', 15' and 7.5' stimuli were respectively N35 - 40.1; 39.9 and 41.3 ms; P50 - 60.5; 64.4 and 65.6 ms and N95 - 103.4; 104.6 and 104.6 ms. For amplitude the normative values for N35-P50 and P50-N95 for 60', 15' and 7.5' were respectively N35-P50 - 1.7; 1.6 and 0.9 µV; P50-N95 - 3.8; 2.8 and 1.5 µV. No gender differences were found either for latency or for amplitude in the three stimulus sizes. There was no correlation between PERG latency and amplitude with age, except for P50 amplitude for stimulus 15'(r=0.39; P=0.035).

Conclusions:

Normative values were determined for PERG parameters of amplitude and latency for three stimulus sizes. These parameters are...
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Reaction Time / Retinal Ganglion Cells / Electroretinography Type of study: Practice guideline Limits: Adolescent / Adult / Female / Humans / Male Language: Portuguese Journal: Arq. bras. oftalmol Journal subject: Ophthalmology Year: 2010 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Reaction Time / Retinal Ganglion Cells / Electroretinography Type of study: Practice guideline Limits: Adolescent / Adult / Female / Humans / Male Language: Portuguese Journal: Arq. bras. oftalmol Journal subject: Ophthalmology Year: 2010 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de São Paulo/BR