Your browser doesn't support javascript.
loading
Próteses fixas implanto-suportadas de 3 elementos: zircônia versus metalocerâmica / Three-unit implant-supported fixed parcial dentures: zirconia versus metal-ceramic
Bonfante, Estevam Augusto; Coelho, Paulo Guilherme; Bonfante, Gerson; Navarro, José Manuel; Thompson, Van Purdy; Pegoraro, Luiz Fernando; Silva, Nelson Renato Franco Alves da.
  • Bonfante, Estevam Augusto; Universidade de São Paulo. Faculdade de Odontologia de Bauru. Bauru. BR
  • Coelho, Paulo Guilherme; New York University. College of Dentistry. New York. US
  • Bonfante, Gerson; Universidade de São Paulo. Faculdade de Odontologia de Bauru. Bauru. BR
  • Navarro, José Manuel; New York University. College of Dentistry. New York. US
  • Thompson, Van Purdy; New York University. College of Dentistry. New York. US
  • Pegoraro, Luiz Fernando; Universidade de São Paulo. Faculdade de Odontologia de Bauru. Bauru. BR
  • Silva, Nelson Renato Franco Alves da; New York University. College of Dentistry. New York. US
Article in Portuguese | LILACS, BBO | ID: lil-561108
RESUMO
A falha coesiva na porcelana de revestimento de próteses com infraestruturas de policristais de zircônia tetragonal estabilizada por ítrio (PZT-I) tem causado elevadas taxas clínicas de fracassos. No presente estudo avaliou-se a confiabilidade e modos de fratura de próteses fixas implanto-suportadas confeccionadas em uma liga de paládio-prata (PdAg) e em PZT-I, submetidas à fadiga acelerada progressiva. Embebeu-se análogos de implantes em uma resina acrílica. Infraestruturas em PZT-I e PdAg (n = 21/grupo) foram fabricadas, revestidas com porcelana, cimentadas e submetidas a uma carga estática (n = 3/grupo) para obtenção dos valores de resistência à fratura. Em seguida, 18 espécimes de cada grupo foram distribuídos ao longo de 3 perfis de estresse progressivo e submetidos à fadiga em água com a carga aplicada na cúspide vestibular do pôntico. Encontrou-se no grupo PdAg um modo de fratura incluindo trincas flexurais na área do conector competindo com a falha coesiva da porcelana, enquanto que as próteses em PZT-I apresentaram falha coesiva na porcelana de revestimento. O cálculo da confiabilidade (Alta Pro, Relia Soft, intervalo de confiança de 90%, bilateral) revelou confiabilidade significativamente inferior para o grupo PZT-I na finalização de uma missão de 100.000 ciclos a 300 N quando comparada ao grupo PdAg. Concluiu-se que para uma missão de 100.000 ciclos a 300 N a confiabilidade é mais alta para a prótese metalocerâmica quando comparada à em PZT-I. Os modos de fratura diferiram entre os grupos.
ABSTRACT
Statement of

problem:

chipping within veneering porcelain has resulted in high clinical failure rates for implant-supported zirconia (Y-TZP) bridges. This study evaluated the reliability and failure modes of mouth-motion step-stress fatigued implant-supported Y-TZP versus Palladium-Silver alloy (PdAg) three-unit bridges.

Methods:

Implant-abutment replicas wereembedded in polymethylmethacrylate resin. Y-TZP and PdAg frameworks, of similar design (n = 21 each), were fabricated, veneered, cemented (n = 3 each), and Hertzian contact tested to obtain ultimate failure load. In each framework group eighteen specimens were distributed across 3 stepstress profiles and mouth-motion cyclically loaded according to the profile on the lingual slope of the buccal cusp of the pontic.

Results:

PdAg failures included competing flexural cracking at abutment and/or connector area and chipping, whereas Y-TZP presented predominantly cohesive failure within veneering porcelain. Including all failure modes, the reliability (2-sided at 90% confidence intervals) for a “mission” of 100,000 cycles at 300 N load was determined (Alta Pro, Reliasoft). Reliability remained unchanged for a mission of 100,000 for PdAg, but significantly decreased for Y-TZP.

Conclusion:

Higher reliability was found for PdAg for a mission of 100,000 cycles at 300N. Failure modes differed between materials.
Subject(s)

Search on Google
Index: LILACS (Americas) Main subject: Zirconium / Ceramics / Dental Prosthesis Language: Portuguese Journal: Innov. implant. j., biomater. esthet. (Impr.) Journal subject: General Surgery / Dentistry / Orthodontics Year: 2009 Type: Article Affiliation country: Brazil / United States Institution/Affiliation country: New York University/US / Universidade de São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: LILACS (Americas) Main subject: Zirconium / Ceramics / Dental Prosthesis Language: Portuguese Journal: Innov. implant. j., biomater. esthet. (Impr.) Journal subject: General Surgery / Dentistry / Orthodontics Year: 2009 Type: Article Affiliation country: Brazil / United States Institution/Affiliation country: New York University/US / Universidade de São Paulo/BR