Avaliação da pressão intraocular através do ocular response analyzer antes e após cirurgia de facoemulsificação / Assessment of intraocular pressure through the ocular response analyzer before and after phacoemulsification surgery
Valbon, Bruno de Freitas; Silva, Renata Siqueira; Jardim, Daniela; Canedo, Ana Laura; Palis, Marcelo; Ambrósio Junior, Renato.
  • Valbon, Bruno de Freitas; Universidade Federal Fluminense. Rio de Janeiro. BR
  • Silva, Renata Siqueira; Fellowship em Glaucoma no Wills Eye Hospital. Philadelphia. US
  • Jardim, Daniela; Instituto de Olhos Renato Ambrósio. Rio de Janeiro. BR
  • Canedo, Ana Laura; Instituto de Olhos Renato Ambrósio. Rio de Janeiro. BR
  • Palis, Marcelo; Universidade Federal Fluminense. Faculdade de Medicina. Rio de Janeiro. BR
  • Ambrósio Junior, Renato; s.af
Rev. bras. oftalmol ; 70(1): 11-15, jan.-fev. 2011. ilus, graf, tab
Article in Portuguese | LILACS | ID: lil-581587
RESUMO

OBJETIVO:

Verificar a variação das medidas pressóricas através do Ocular Response Analyzer (ORA, Reichert) IOPg (gold standard, Goldmann correlated intraocular pressure) e IOPcc (corneal compensated intraocular pressure) antes e após a cirurgia de catarata por facoemulsificação.

MÉTODOS:

Estudo observacional prospectivo envolvendo 47 olhos de 36 pacientes que se submeteram à cirurgia de catarata por facoemulsificação com implante de lente intraocular. A Pressão intraocular compensada da córnea (IOPcc) e correlacionada com Goldmann (IOPg), foram medidos pelo Ocular Response Analyzer (ORA); As medidas foram realizadas no pré-operatório, 1º dia e 30º dia após a cirurgia de facoemulsificação. O Teste de Kolmogorov-Smirnov foi utilizado para avaliar a distribuição normal. Diferenças obtidas entre as medidas obtidas através do Ocular Response Analyzer foram avaliadas pelo T-test (Student's paired t-test). Foi considerado como estatisticamente significante p < 0,05 por cento.

RESULTADOS:

As médias da IOPcc no pré-operatório, no primeiro dia e no trigésimo dia após a cirurgia de facoemulsificação e implante de LIO foram de 15.6 (±3.06), 15.5 (±3.14) , e 15.71 (±4.34). A média de IOPg no pré-operatório foi de 14.7 (± 3.0), no primeiro dia após cirurgia 12.9 (± 3.1) e no trigésimo dia 13.4 (± 3.7). Houve diferença significativa entre IOPg (PRE X D1) com p < 0,01.

CONCLUSÃO:

Observamos uma diminuição da IOPg no pós-operatório imediato em córneas com edema após facoemulsificação. Logo, ressaltamos a importância da biomecânica na medida da Pio.
ABSTRACT

PURPOSE:

To evaluate the IOPg (gold standard, Goldmann correlated Intraocular pressure) and IOPcc (corneal compensated Intraocular pressure) in patients prior to and following phacoemulsification with intraocular lens implantation

METHODS:

This is a prospective study. Thirty six consecutive patients (47 eyes) that underwent phacoemulsification surgery and intraocular lens implantation were enrolled. Full ophthalmologic clinical assessment including Ocular Response Analyzer (ORA, Reichert Ophthalmic Instruments, Buffalo, NY, USA) measurements were performed in all cases prior to surgery (Pre) and at follow-up time points day 1 and day 30. Exclusion criteria were corneal disease or scars, previous ocular surgery and patients with glaucoma or glaucoma suspect or ocular hypertension. Student's paired t-test was utilized.

RESULTS:

Mean and standard-deviation of ORA parameters were IOPcc Pre was 15.6 (±3.06), day 1 (D1) 15.5 (±3.14) and day 30 (D30) 15.71 (±4.34). The mean of IOPg was Pre 14.7 (± 3.0), day 1 (D1) 12.9 (± 3.1), and day 30 (D30) was 13.4 (± 3.7).

CONCLUSION:

The reduction in the IOPg was observed in the immediate postoperative underestimating the real intraocular pressure ocular. We agree that biomechanical properties may have greater impact on IOP measurements error than corneal tchikness or curvature.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Phacoemulsification / Cornea / Lens Implantation, Intraocular / Diagnostic Techniques, Ophthalmological / Intraocular Pressure Type of study: Observational study Limits: Humans Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2011 Type: Article Affiliation country: Brazil / United States Institution/Affiliation country: Fellowship em Glaucoma no Wills Eye Hospital/US / Instituto de Olhos Renato Ambrósio/BR / Universidade Federal Fluminense/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Phacoemulsification / Cornea / Lens Implantation, Intraocular / Diagnostic Techniques, Ophthalmological / Intraocular Pressure Type of study: Observational study Limits: Humans Language: Portuguese Journal: Rev. bras. oftalmol Journal subject: Ophthalmology Year: 2011 Type: Article Affiliation country: Brazil / United States Institution/Affiliation country: Fellowship em Glaucoma no Wills Eye Hospital/US / Instituto de Olhos Renato Ambrósio/BR / Universidade Federal Fluminense/BR