Validación del método analítico para la determinación de valsartán en plasma humano por HPLC/UV con adición de estándar empleando losartán como estándar interno / Validation of an analytical method for the determination of valsartan in human plasma by HPLC/UV with addition standard using losartan as an internal standard
Pérez, Milena; Ramírez, Gloria; Pérez, Mauricio; Restrepo, Piedad.
  • Pérez, Milena; CIDEIM. Laboratorio Biofarmacéutico. Santiago de Cali. CO
  • Ramírez, Gloria; CIDEIM. Santiago de Cali. CO
  • Pérez, Mauricio; CIDEIM. Unidad de Epidemiología y Bioestadística. Santiago de Cali. CO
  • Restrepo, Piedad; CIDEIM. Unidad de Estudios Biofarmacéuticos. Santiago de Cali. CO
Colomb. med ; 38(1): 13-20, ene.-mar. 2007. ilus, tab
Article in Spanish | LILACS | ID: lil-586337
RESUMEN
Introducción: El concepto de validación se refiere a la evaluación estadística de los resultados obtenidos en la aplicación de técnicas analíticas, mediante pruebas convenientemente documentadas y demostrativas de que un método es lo suficientemente fiable a fin de producir el resultado previsto bajo condiciones definidas, como son: sistema analítico, intervalo de concentración, infraestructura y talento humano. Objetivo: Describir el proceso de validación del método analítico para la cuantificación de valsartán en plasma humano por HPLC-UV y su aplicación en estudios farmacocinéticos, de biodisponibilidad y bioequivalencia de medicamentos que contengan valsartán como principio activo. Metodología: Se desarrolló un método para la detección y cuantificación de valsartán en plasma humano, con elución isocrática por cromatografía líquida de fase reversa, con detección ultravioleta a 265 nm, mediante el método de adición de estándar. Se utilizó losartán como estándar interno. El método implica una extracción en fase sólida de los principios activos (valsartán y losartán) con cartuchos C8. La separación se realizó en una columna C18 en fase reversa y la fase móvil fue una mezcla de acetonitrilo: buffer fosfato (45:55 v/v) ajustado a pH 2.7±0.1 con ácido fosfórico concentrado. El método se validó en el rango de concentraciones de 0.05 a 20 µg/ml con adición de estándar de valsartán de 2.5 µg/ml. Resultados y conclusiones: La curva de calibración fue lineal en el rango de concentraciones establecido. Se evaluó la reproducibilidad, estabilidad y porcentaje de recuperación del método. El método para determinar el valsartán en plasma humano por HPLC/UV fue preciso y exacto con un límite de cuantificación de 1.485 µg/ml. Este método fue suficientemente sensible para su aplicación en estudios farmacocinéticos de valsartán.
ABSTRACT
Introduction: The validation concept refers to the statistical evaluation of the results obtained in the application of analytic technics, by appropriately documented and demonstrative tests that a method is sufficiently reliable to produce the result foreseen under defined conditions, like they are: analytic system, concentration interval, infrastructure and human talent.Objective: To describe the validation process of the analytic method for the valsartan quantification in human plasma by HPLC-UV and its application in pharmacokinetic, bioavailability and bioequivalence studies of products that contain the active principle valsartan.Methodology: A method for detection and quantification of valsartan in human plasma has been developed using an isocratic elution on reversed phase liquid chromatography with ultraviolet detection at a single wavelength (265 nm) and the addition standard method. Losartan was used as an internal standard. This method involves a solid-phase drug extraction (valsartan and losartan) from plasma using C8 cartridges. Separation was achieved on a C18 reversed phase column and the mobile phase consisted of 45% acetonitrile and 55% phosphate buffer (adjusted to pH 2.7 ± 0.1 with phosphoric acid). The assay has been vali-dated over a concentration range of 0.05 to 20 µg/ml with addition of valsartan 2.5 µg/ml. Results and conclusions: Calibration curve was linear in the described concentration range. The reproducibility, stability and recovery of the method were evaluated. Determination of valsartan in human plasma by HPLC/UV method was accurate and precise with a quantitation limit of 1.485 µg/ml. The method was sufficiently sensitive for pharmacokinetic studies of valsartan in human plasma.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Laboratory and Fieldwork Analytical Methods / Chromatography, High Pressure Liquid / Losartan / Solid Phase Extraction Limits: Humans Language: Spanish Journal: Colomb. med Journal subject: Medicine Year: 2007 Type: Article Affiliation country: Colombia Institution/Affiliation country: CIDEIM/CO

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Laboratory and Fieldwork Analytical Methods / Chromatography, High Pressure Liquid / Losartan / Solid Phase Extraction Limits: Humans Language: Spanish Journal: Colomb. med Journal subject: Medicine Year: 2007 Type: Article Affiliation country: Colombia Institution/Affiliation country: CIDEIM/CO