Avaliação pela tomografia de coerência óptica de stent nacional recoberto com polímero biodegradável eluidor de sirolimus vs. stent eluidor de biolimus A9 em artérias coronárias porcinas / Assessment by optical coherence tomography of a Brazilian biodegradable polymer-coated sirolimus eluting stent vs a biolimus A9-eluting stent in porcine coronary arteries
Takimura, Celso Kiyochi; Galon, Micheli Zanotti; Lopes Junior, Augusto Celso de Araújo; Carvalho, Juliana; Ferreira, Suzane Kiss; Chaves, Márcio José Figueira; Aiello, Vera Demarchi; Gutierrez, Paulo Sampaio; Laurindo, Francisco Rafael Martins; Lemos, Pedro Alves.
  • Takimura, Celso Kiyochi; USP. FM. Instituto do Coração. São Paulo. BR
  • Galon, Micheli Zanotti; USP. FM. Instituto do Coração. São Paulo. BR
  • Lopes Junior, Augusto Celso de Araújo; USP. FM. Instituto do Coração. São Paulo. BR
  • Carvalho, Juliana; USP. FM. Instituto do Coração. São Paulo. BR
  • Ferreira, Suzane Kiss; USP. FM. Instituto do Coração. São Paulo. BR
  • Chaves, Márcio José Figueira; USP. FM. Instituto do Coração. São Paulo. BR
  • Aiello, Vera Demarchi; USP. FM. Instituto do Coração. São Paulo. BR
  • Gutierrez, Paulo Sampaio; USP. FM. Instituto do Coração. São Paulo. BR
  • Laurindo, Francisco Rafael Martins; USP. FM. Instituto do Coração. São Paulo. BR
  • Lemos, Pedro Alves; USP. FM. Instituto do Coração. São Paulo. BR
Rev. bras. cardiol. invasiva ; 19(2): 138-144, jul. 2011. tab, ilus
Article in Portuguese | LILACS | ID: lil-595226
RESUMO

Introdução:

Está em desenvolvimento em nosso meio o stent InspironTM, um stent nacional metálico de cobalto-cromorecoberto com uma mistura de polímeros bioabsorvíveis exclusivamente em sua face abluminal e com liberação de sirolimus. O objetivo deste estudo foi comparar, em um modelo experimental, os achados da tomografia de coerência óptica dos stents InspironTM e BioMatrixTM.

Métodos:

Os stents foram implantados em artérias coronárias porcinas. Cada indivíduo recebeu os dois tipos de stent, um em cada artéria (artérias descendente anterior e circunflexa), e após 28 dias foi realizado novo estudo angiográfico e empregada a tomografia de coerência óptica para avaliação da hiperplasia neointimal.

Resultados:

Sete porcos receberam 7 stents InspironTM e 7 stents BioMatrixTM. O diâmetro de referência dos vasos tratados foi de 2,16 + 0,37 mm e a relação balão de implante do stent/artéria foi de 1,17 + 0,16 atm, sem diferença entre os grupos. A perda tardia intrastent foi de 0,53 + 0,56 mm e de 0,32 + 0,37 mm (P = 0,43) para os stents InspironTM e BioMatrixTM, respectivamente. Ocorreu um caso de reestenose angiográfica no grupo stent InspironTM (14% vs. 0; P = 0,2). A tomografia de coerência óptica evidenciou área neointimal intrastent no stent InspironTM de 1,61 + 0,57 mm² e no stent BioMatrixTM de 1,36 + 0,66 mm² (P = 0,47). A área neointimal porcentual foi de 31% no stent InspironTM vs. 23% no stent BioMatrixTM (P = 0,21).

Conclusões:

Após 28 dias de implante em artérias coronárias porcinas o stent InspironTM apresentou grau de hiperplasia neointimal intrastent semelhante ao do stent BioMatrixTM.
ABSTRACT

Background:

InspironTM, a chrome-cobalt sirolimus-eluting stent covered by a mixture of bioabsorbable polymers on its abluminal side is being developed in Brazil. Our objective was to compare in an experimental study the findings ofoptical coherence tomography for the InspironTM and BioMatrix™ stents.

Methods:

Stents were implanted in porcine coronary arteries. Each individual received two types of stents, one in each artery (left anterior descending and left circumflex artery). After 28 days, a new angiography was performed and optical coherence tomography was used to assess neointimal hyperplasia.

Results:

Seven Inspiron™ stents and 7 BioMatrixTM stents were implanted in 7 domestic pigs. The reference diameter of the treated vessels was 2.16 + 0.37 mm and the balloon to artery ratio was 1.17 + 0.16 atm, with no statisticaldifference between groups. In-stent late loss was 0.53 + 0.56 mm and 0.32 + 0.37 mm (P = 0.43) for the InspironTMand BioMatrixTM stents, respectively. There was one case of angiographic restenosis in the InspironTM group (14% vs 0; P = 0.2). Optical coherence tomography showed an instent neointimal area of 1.61 + 0.57 mm² for the InspironTM stent and 1.36 + 0.66 mm² for the BioMatrixTM stent (P = 0.47). The percentage of neointimal area was 31% for the InspironTM stent vs 23% for the BiomatrixTM stent (P = 0.21).

Conclusions:

Twenty-eight days after implantation in porcine coronaryarteries the InspironTM stent showed a similar degree of neointimal hyperplasia as the BiomatrixTM stent.
Subject(s)
Full text: Available Index: LILACS (Americas) Main subject: Tomography, Optical Coherence / Drug-Eluting Stents Type of study: Prognostic study Limits: Animals Country/Region as subject: South America / Brazil Language: Portuguese Journal: Rev. bras. cardiol. invasiva Journal subject: Cardiology / General Surgery Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: USP/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Tomography, Optical Coherence / Drug-Eluting Stents Type of study: Prognostic study Limits: Animals Country/Region as subject: South America / Brazil Language: Portuguese Journal: Rev. bras. cardiol. invasiva Journal subject: Cardiology / General Surgery Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: USP/BR