Os perfusionistas Brasileiros e o ajuste do rolete arterial: comparação entre a calibração estática e dinâmica / Brazilian perfusionists and arterial roller pump adjustment: comparison between static and dynamic calibration method
Vieira Junior, Francisco Ubaldo; Antunes, Nilson; Medeiros Júnior, Johannes Dantas de; Vieira, Reinaldo Wilson; Carvalho Filho, Élio Barreto de; Reis Junior, José Evaldo Cavalcante; Costa, Eduardo Tavares.
  • Vieira Junior, Francisco Ubaldo; s.af
  • Antunes, Nilson; Universidade Estadual de Campinas. Hospital das Clínicas. Campinas. BR
  • Medeiros Júnior, Johannes Dantas de; UNICAMP. FEEC. Departamento de Engenharia Biomédica. Campinas. BR
  • Vieira, Reinaldo Wilson; Universidade Estadual de Campinas. Faculdade de Ciências Médicas. Campinas. BR
  • Carvalho Filho, Élio Barreto de; Universidade Estadual de Campinas. Hospital das Clínicas. Campinas. BR
  • Reis Junior, José Evaldo Cavalcante; Universidade Estadual de Campinas. Hospital das Clínicas. Campinas. BR
  • Costa, Eduardo Tavares; UNICAMP. Centro de Engenharia Biomédica. Campinas. BR
Rev. bras. cir. cardiovasc ; 26(2): 205-212, abr.-jun. 2011. ilus, tab
Article in Portuguese | LILACS | ID: lil-597740
RESUMO

INTRODUÇÃO:

Bombas de roletes desempenham um papel importante na circulação extracorpórea. No entanto, a oclusão dos roletes das bombas deve ser realizada de forma adequada e estas podem ser ajustadas, principalmente por dois

métodos:

estático e dinâmico.

OBJETIVO:

Investigar como os perfusionistas brasileiros ajustam as bombas de rolete arterial em seus serviços e testar o uso de um Dispositivo Auxiliar de Calibração que facilita o ajuste pelo método de calibração dinâmica.

MÉTODOS:

Foi instalada uma bomba de roletes com os acessórios necessários para a realização de sua calibração pelos métodos de velocidade de queda (calibração estática) e calibração dinâmica durante o XXVIII Congresso Brasileiro de Circulação Extracorpórea. Foi solicitado aos perfusionistas que ajustassem uma bomba de roletes conforme procedimento normalmente utilizado em seu serviço. Após cada regulagem, foi medida a respectiva pressão pelo método de calibração dinâmica com o auxílio do dispositivo. O projeto foi aprovado pelo Comitê de Ética em Pesquisa da UNICAMP, sob Nº 1144/2010.

RESULTADOS:

Participaram da pesquisa 56 perfusionistas. A média das 56 medidas de pressão de calibração dinâmica foi 434 ± 214 mmHg; 76 por cento das medidas realizadas ficaram no intervalo preconizado para o uso da técnica de calibração dinâmica (entre 150 e 500 mmHg).

CONCLUSÃO:

Os perfusionistas brasileiros tendem a calibrar bombas de roletes com ajustes menos oclusivos. As amplitudes das medidas de pressão de calibração dinâmica tendem a ser menores para perfusionistas mais experientes. O dispositivo pode ser utilizado por perfusionistas para ajustar bombas de roletes com maior precisão e, principalmente, repetitividade e em alguns minutos.
ABSTRACT

INTRODUCTION:

Roller pumps play an important role in extracorporeal circulation. However, occlusion of the rollers should be adequately performed and this can be adjusted mainly by two

methods:

static and dynamic.

OBJECTIVE:

To investigate how the Brazilian perfusionists adjust arterial roller pumps in their services and evaluate the application of a Device to Assist Calibration (DAC) that facilitates roller adjustment by the dynamic calibration method.

METHODS:

We installed a roller pump with accessories to perform adjustment by drop rate (static calibration) and dynamic calibration methods during the XXVIII Brazilian Congress of Extracorporeal Circulation. Perfusionists were asked to adjust the roller pump according to the procedure they usually do in their service. After each adjustment pressure was measured by dynamic calibration method with DAC. The research was approved by the Research Ethics Committee of UNICAMP, Nº 1144/2010.

RESULTS:

There were 56 perfusionists in this study. Pressure average of 56 measurements of dynamic calibration was 434 ± 214 mmHg; 76 percent of measurements were within the recommended range for the use of the dynamic calibration method (between 150 and 500 mmHg).

CONCLUSION:

Brazilian perfusionists tend to adjust roller pumps with less occlusive settings. The amplitudes of the dynamic calibration pressure tend to be smaller for more experienced perfusionists because their skills increase with time. The device can be used by the perfusionist to adjust roller pumps with greater accuracy and mainly repeatability in few minutes.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Infusion Pumps / Extracorporeal Circulation Type of study: Prognostic study Limits: Adult / Female / Humans / Male Country/Region as subject: South America / Brazil Language: Portuguese Journal: Rev. bras. cir. cardiovasc Journal subject: Cardiology / General Surgery Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: UNICAMP/BR / Universidade Estadual de Campinas/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Infusion Pumps / Extracorporeal Circulation Type of study: Prognostic study Limits: Adult / Female / Humans / Male Country/Region as subject: South America / Brazil Language: Portuguese Journal: Rev. bras. cir. cardiovasc Journal subject: Cardiology / General Surgery Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: UNICAMP/BR / Universidade Estadual de Campinas/BR