Comparação e correlação entre automedida, medida casual e monitorização ambulatorial da pressão arterial / Comparison and correlation between self-measured blood pressure, casual blood pressure measurement and ambulatory blood pressure monitoring / Comparación y correlación entre automedida, medida casual y monitorización ambulatoria de la presión arterial
Souza, Weimar Kunz Sebba Barroso de; Jardim, Paulo César Brandão Veiga; Porto, Ludmila Brito; Araújo, Fabrício Alves; Sousa, Ana Luíza Lima; Salgado, Cláudia Maria.
  • Souza, Weimar Kunz Sebba Barroso de; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
  • Jardim, Paulo César Brandão Veiga; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
  • Porto, Ludmila Brito; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
  • Araújo, Fabrício Alves; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
  • Sousa, Ana Luíza Lima; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
  • Salgado, Cláudia Maria; Universidade Federal de Goiás. Faculdade de Medicina. Liga de Hipertensão Arterial. Goiânia. BR
Arq. bras. cardiol ; 97(2): 148-155, ago. 2011. ilus, tab
Article in Portuguese | LILACS | ID: lil-601780
RESUMO
FUNDAMENTO A medida casual da pressão arterial (PA) pelos profissionais de saúde está sujeita a uma grande variabilidade, sendo necessário buscar novos métodos que possam superar essa limitação.

OBJETIVO:

Comparar e avaliar a correlação entre os níveis de PA obtidos por meio da automedida da pressão arterial (AMPA) com a medida casual e com a monitorização ambulatorial da pressão arterial (MAPA).

MÉTODOS:

Avaliamos hipertensos que realizaram as três metodologias de medida da PA com intervalo menor que 30 dias; as médias das pressões foram utilizadas para comparação e correlação. Foram empregados os aparelhos OMRON 705 CP (medida casual), OMRON HEM 714 (AMPA) e SPACELABS 9002 (MAPA).

RESULTADOS:

Foram avaliados 32 pacientes, 50,09 por cento mulheres, idade média 59,7 (± 11,2) anos, média do IMC 26,04 (± 3,3) kg/m². Valores médios de pressão sistólica (PAS) e pressão diastólica (PAD) para a AMPA foram de 134 (± 15,71)mmHg e 79,32 (± 12,38) mmHg. Na medida casual as médias da PAS e PAD foram, respectivamente, 140,84 (± 16,15)mmHg e 85 (± 9,68) mmHg. Os valores médios da MAPA na vigília foram 130,47 (± 13,26) mmHg e 79,84 (± 9,82) mmHg para PAS e PAD, respectivamente. Na análise comparativa, a AMPA apresentou valores semelhantes aos da MAPA (p > 0,05) e diferentes da medida casual (p < 0,05). Na análise de correlação a AMPA foi superior à medida casual, considerando a MAPA como o padrão de referência nas medidas tensionais.

CONCLUSÃO:

A AMPA apresentou melhor comparação com a MAPA do que a medida casual e também se correlacionou melhor com a aquela, especialmente para a pressão diastólica, devendo ser considerada uma alternativa com baixo custo para o acompanhamento do paciente hipertenso.
ABSTRACT

BACKGROUND:

Casual blood pressure (BP) measurement by healthcare professionals is subject to great variability and new methods are necessary to overcome this limitation.

OBJECTIVE:

To compare and assess the correlation between the BP levels obtained by self-measured BP (SMBP), casual BP measurement and ambulatory blood pressure monitoring (ABPM).

METHODS:

We assessed hypertensive individuals submitted to the three methods of BP measurement at an interval < 30 days; the BP means were used for comparison and correlation. The following devices were used OMRON 705 CP (casual measurement), OMRON HEM 714 (SMBP) and SPACELABS 9002 (ABPM).

RESULTS:

A total of 32 patients were assessed, of which 50.09 percent were females, with a mean age of 59.7 (± 11.2), BMI mean of 26.04 (± 3.3) kg/m². Mean systolic (SBP) and diastolic blood pressure (DBP) for SMBP were 134 (± 15.71) mmHg and 79.32 (± 12.38) mmHg. The casual measurement means of SBP and DBP were, respectively, 140.84 (± 16.15) mmHg and 85 (± 9.68) mmHg. The mean values of ABPM during the wakefulness period were 130.47 (± 13.26) mmHg and 79.84 (± 9.82) mmHg for SBP and DBP, respectively. At the comparative analysis, the SMBP had similar results to those obtained at ABPM (p > 0.05) and different from the casual measurement (p < 0.05). At the analysis of correlation, SMBP values were higher than the casual measurements, considering ABPM as the reference standard in BP measurements.

CONCLUSION:

SMBP showed a better correlation with ABPM than the casual measurement and was also better correlated with the latter, especially regarding the DBP and should be considered as a low-cost alternative for the follow-up of the hypertensive patient.
RESUMEN
FUNDAMENTO La medida casual de la presión arterial (PA), por los profesionales de la salud está sujeta a una gran variabilidad, siendo necesario buscar nuevos métodos que puedan superar esa limitación.

OBJETIVO:

Comparar y evaluar la correlación entre los niveles de PA obtenidos por medio de la automedida de la presión arterial (AMPA) con la medida casual y con la monitorización ambulatoria de la presión arterial (MAPA).

MÉTODOS:

Evaluamos hipertensos que realizaron las tres metodologías de medida de la PA con un intervalo menor que 30 días. Los promedios de las presiones fueron utilizados para la comparación y la correlación. Fueron usados los aparatos OMRON 705 CP (medida casual), OMRON HEM 714 (AMPA) y SPACELABS 9002 (MAPA).

RESULTADOS:

Se evaluaron 32 pacientes, 50,09 por ciento mujeres, edad promedio 59,7(± 11,2) años, promedio del IMC 26,04 (± 3,3) kg/m². Los valores promedios de presión sistólica (PAS) y presión diastólica (PAD) para la AMPA fueron de 134 (± 15,71) mmHg y 79,32(± 12,38) mmHg. En la medida casual los promedios de la PAS y PAD fueron, respectivamente, 140,84 (± 16,15) mmHg y 85 (± 9,68) mmHg. Los valores promedios de la MAPA en la vigilia fueron 130,47 (± 13,26) mmHg y 79,84 (± 9,82) mmHg para PAS y PAD, respectivamente. En el análisis comparativo, la AMPA tuvo valores similares a los de la MAPA (p > 0,05) y diferentes de la medida casual (p < 0,05). En el análisis de correlación la AMPA fue superior a la medida casual, considerando la MAPA como el estándar de referencia en las medidas tensionales.

CONCLUSIÓN:

La AMPA presentó una mejor comparación con la MAPA que la medida casual y también se correlacionó mejor con aquella, especialmente para la presión diastólica, debiendo ser considerada como una alternativa con bajo coste para el seguimiento del paciente hipertenso.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Blood Pressure / Blood Pressure Determination / Hypertension Type of study: Observational study Limits: Adult / Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de Goiás/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Blood Pressure / Blood Pressure Determination / Hypertension Type of study: Observational study Limits: Adult / Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2011 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de Goiás/BR