Your browser doesn't support javascript.
loading
Comparação da eficácia dos modos de desmames de ventilação mecânica automatizados: um estudo de bancada / Comparison of the effectiveness of modes of automated weaning from mechanical ventilation: a bench study
São Paulo; s.n; 2011. [163] p. ilus, tab, graf.
Thesis in Portuguese | LILACS | ID: lil-609375
RESUMO

INTRODUÇÃO:

O desmame da ventilação mecânica é um processo complexo que requer avaliação e interpretação de parâmetros clínicos objetivos e subjetivos. O atraso no processo de desmame pode expor o paciente a um desconforto desnecessário, aumentar o risco de complicações e custos. Os modos de desmame automatizados podem acelerar a extubação e diminuir a carga de trabalho da equipe da UTI. Há diversos modos automatizados de desmame disponíveis que foram avaliados, separadamente, em populações selecionadas, com resultados divergentes em relação ao desmame convencional. No entanto, os modos automáticos desmame não foram comparados entre si, nem sistematicamente avaliados, em condições específicas, mas comuns, como a ansiedade extrema ou esforços inspiratórios inefetivos.

OBJETIVOS:

Comparar os modos Smartcare®, ASV® e MRV® quanto a eficácia no desmame da ventilação mecânica.

MÉTODOS:

Estudo de bancada para avaliar os três diferentes modos de desmame automatizado adaptive support ventilation (ASV®), mandatory rate ventilation (MRV®) and Smartcare®. Nós simulamos os pacientes usando um simulador pulmonar programável (ASL 5000 - Ingmar Medical) com o padrão respiratório, mecânica respiratória e CO2 arterial de derivados de artigos publicados em periódicos médicos para criar duas condições 1.extubação provável mecânica pulmonar normal, ansiedade extrema, idosos normais, padrão de respiração irregular extrema (Cheyne-Stokes), doença pulmonar restritiva; 2. extubação improvável mecânica pulmonar alterada, com e sem esforços inspiratórios ineficazes.

RESULTADOS:

Os pacientes com extubação possível, ansiedade extrema, o padrão de respiração irregular moderada e extubação impossível foram diagnosticados corretamente por todos os modos. Os pacientes com Cheyne-Stokes foram diagnosticados impropiamennte por todos os modos, mas o modo Smartcare® diagnosticou corretamente quando a opção de distúrbio neurológico foi ativado. Apenas o Smartcare® diagnosticou...
ABSTRACT

INTRODUCTION:

Weaning from mechanical ventilation is a complex process requiring assessment and interpretation of both objective and subjective clinical parameters. Delay in weaning process may expose the patient to unnecessary discomfort and increased risk of complications, and increasing the cost of care. Automated weaning modes could quicken the extubation and decrease the ICU team workload. Many automated weaning modes were now available. They were separately evaluated in selected populations with divergent results when compared to conventional weaning. However the automated weaning modes were not compared among them, neither systematically evaluated in challenging but common conditions, as extreme anxiety or ineffective inspiratory efforts.

OBJECTIVES:

Compare Smartcare®, ASV® and MRV® effectiveness in weaning of mechanical ventilation.

METHODS:

Bench study to evaluate three different automated weaning modes adaptive support ventilation (ASV®), mandatory rate ventilation (MRV®) and Smartcare®. We simulated the patients using a programmable lung simulator (ASL 5000 Ingmar Medical) with the breathing pattern, respiratory mechanics and arterial CO2 derived from published medical journals articles to create two conditions 1. Successful extubation normal lung mechanic, extreme anxiety; old normal adult, extreme irregular breathing pattern (Cheyne-Stokes), restrictive lung disease; 2. Unsuccessful extubation altered pulmonary mechanics with and without ineffective inspiratory efforts.

RESULTS:

Patients with successful extubation, extreme anxiety, moderate irregular breathing pattern and unsuccessful extubation were properly diagnosed by all modes. Patients with Cheyne-Stokes were improperly diagnosed by all modes, but the Smartcare® mode properly diagnosed when the neurologic disorder option was activated. Only Smartcare® properly diagnosed the patient with rapid shallow breathing due to restrictive lung disease. Only MRV improperly diagnosed the...
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Respiration, Artificial / Respiratory Therapy / Ventilators, Mechanical Language: Portuguese Year: 2011 Type: Thesis

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Respiration, Artificial / Respiratory Therapy / Ventilators, Mechanical Language: Portuguese Year: 2011 Type: Thesis