Desarrollo y validación de un método analítico para el control de calidad y estabilidad de fenilefrina 10 por ciento colirio y tropicamida 1 por ciento / Development and validation of an analytical method for quality control and the stability of the eyedrops 10 % Phenylephrine and the 1 % Tropicamide
Rev. cuba. farm
;
45(2): 205-215, Apr.-June 2011.
Article
in Spanish
| LILACS
| ID: lil-615145
RESUMEN
Se desarrolló y validó un método analítico por cromatografía líquida de alta resolución, aplicable al control de la calidad y el estudio de estabilidad de fenilefrina 10 por ciento más tropicamida 1 percent, colirio. Para cuantificar simultáneamente los 2 principios activos en el producto terminado, la separación se realizó a través de una columna cromatográfica Lichrosorb RP-18 (5 µm) (250 x 4 mm), con detección ultravioleta a 253 nm, empleando una fase móvil compuesta por metanolagua destilada (11), con 1,1 g de 1-octanosulfonato de sodio por litro ajustado pH a 3,0 con ácido fosfórico y la cuantificación de este frente a una muestra de referencia con el método del estándar externo. El método analítico desarrollado resultó lineal, preciso, específico y exacto en el rango de concentraciones estudiadas, el cual se estableció para el control de la calidad y el estudio de estabilidad del producto terminado ya que no existían métodos analíticos reportados para estos fines
ABSTRACT
An analytical high-performance liquid chromatography method was developed and validated applicable to quality control and to stability study of 10 percent phenylephrine plus eyedrops 1 percent tropicamide. To quantify simultaneously both active principles in the finished product, separation was carried out through a Lichrosorb RP-18 (15 µm) (260 x 4 mm) column chromatography, with ultraviolet detection at 253 nm using the mobile phase composed of methanol distilled water (11), with 1.1 g of sodium 1-octasulfanate by litre and pH fitted to 3,0 with phosphoric acid and the quantification of this front to a reference sample using the external standard method. The analytical method developed was linear, precise, specific and accurate in the rank of study concentrations, established for the quality control and stability study of the finished product since there were not analytical methods designed for these aims:
Full text:
Available
Index:
LILACS (Americas)
Main subject:
Quality Control
/
Pharmaceutical Preparations
Language:
Spanish
Journal:
Rev. cuba. farm
Journal subject:
Pharmacy
Year:
2011
Type:
Article
Affiliation country:
Cuba
Institution/Affiliation country:
Centro de Investigación y Desarrollo de Medicamentos/CU
/
Empresa Farmacéutica Julio Trigo/CU