Your browser doesn't support javascript.
loading
Avaliação morfológica dos conjuntos guia excêntrico: placas dcp-l 4.5 mm / Morphological evaluation of eccentric sets guide-plates of dcp-l 4.5 mm
Borges, José Humberto de Souza; Silva Filho, Antonio Lisboa da; Pereira Neto, Francisco; Daher, Walter Rodrigo; Mesquita, Alessandro Queiroz de; Freitas, Anderson.
  • Borges, José Humberto de Souza; s.af
  • Silva Filho, Antonio Lisboa da; s.af
  • Pereira Neto, Francisco; s.af
  • Daher, Walter Rodrigo; s.af
  • Mesquita, Alessandro Queiroz de; s.af
  • Freitas, Anderson; s.af
Acta ortop. bras ; 20(1): 13-16, 2012. ilus, tab
Article in Portuguese | LILACS | ID: lil-616920
RESUMO

OBJETIVO:

Avaliar isolada e comparativamente medidas do conjunto guia excêntrico-placa utilizado no material de 4.5mm e determinar sua influência na força de compressão.

MÉTODOS:

Foram utilizados quatros guias excêntricos e quatro placas largas de compressão dinâmica (DCP-L) de quatro fabricantes nacionais e um paquímetro digital Vonder®. Criou-se medidas-padrão para os conjuntos, totalizando cinco parâmetros, identificados de A a E. Os conjuntos foram formados por materiais da mesma fábrica e identificados de I a IV. As análises foram feitas com a visualização volar dos guias inseridos nas placas. As placas foram ensaiadas em máquina universal de ensaios.

RESULTADOS:

Os conjuntos I, II, apresentaram os mesmos valores em todas as medidas. Todos os conjuntos apresentaram os mesmos valores de E = 8.15 e B = 3.60. O conjunto III A= 8.10mm, C=3.25mm e D= 1.25mm. O conj. IV A = 7.00mm, C=3.10mm, D = 0.30mm. O conj. I apresentou Força de compressão máxima (F Max.) 80.58 N, conj. II F Max. 81.63 N, conj. III F Max. 36.32N, conj. IV F Max. 37.52N.

CONCLUSÃO:

As medidas avaliadas nos conjuntos denotam a falta de padronização na fabricação de instrumentais ortopédicos e sua consequência nos valores da força de compressão. Nível de evidência III, estudo analítico.
ABSTRACT

OBJECTIVE:

To carry out isolated and comparative evaluations of the measurements of the set eccentric guide plates used in 4.5mm surgical implants, and to determine the effect of these measurements on compression strength.

METHODS:

Four eccentric guides, (or guide plates?) four large dynamic compression plates (L-DCP) from four local manufacturers, and a Vonder® 200 mm caliper brand were used. Five standard parameter measurements were created for the set eccentric guide-plate, which were identified as A to E. Four sets were made, using materials of the same factory, and identified as groups I to IV. The analyses were performed by measuring all the parameters from a ventral view of the plate, with the eccentric guide placed in the plate hole.

RESULTS:

Groups I and II showed the same values for all the parameters. All the groups showed the same measurements for E = 8.15 e B = 3.60. Group III A = 8.10mm, C = 3.25mm, D = 1.25mm. Group IV A = 7.00mm, C = 3.10mm, D = 0.30mm. Maximum compression force was (F Max.) Group I 80.58 N, Group. II F Max. 81.63 N, Group. III F Max. 36.32N, Group. IV F Max. 37.52N

CONCLUSION:

The measurements evaluated show a lack of standardization in the manufacture of orthopedic instruments and its effects on the values for compression strength. Level of Evidence Level III, analytical study.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Bone Plates / Bone Screws / Fracture Fixation, Internal Type of study: Evaluation studies / Prognostic study Limits: Humans Language: Portuguese Journal: Acta ortop. bras Journal subject: Orthopedics Year: 2012 Type: Article

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Bone Plates / Bone Screws / Fracture Fixation, Internal Type of study: Evaluation studies / Prognostic study Limits: Humans Language: Portuguese Journal: Acta ortop. bras Journal subject: Orthopedics Year: 2012 Type: Article