Your browser doesn't support javascript.
loading
Fones de inserção: aplicação no colabamento de meato acústico externo / Insert earphones: application to avoid collapse of the external auditory canal
Marangoni, Andréa Tortosa; Scharlach, Renata Coelho; Silveira, Marcela Rosana Maia da; Calais, Lucila Leal; Gil, Daniela.
  • Marangoni, Andréa Tortosa; s.af
  • Scharlach, Renata Coelho; Universidade Federal de São Paulo. São Paulo. BR
  • Silveira, Marcela Rosana Maia da; s.af
  • Calais, Lucila Leal; s.af
  • Gil, Daniela; Universidade Federal de São Paulo. São Paulo. BR
Rev. Soc. Bras. Fonoaudiol ; 17(1): 61-65, jan.-mar. 2012. ilus, tab
Article in Portuguese | LILACS | ID: lil-617221
RESUMO

OBJETIVO:

Determinar a eficácia dos fones de inserção em casos de suspeita de colabamento de meato acústico externo.

MÉTODOS:

Vinte e um idosos com intervalo aéreo-ósseo superior a 10 dBNA nas frequências de 3 kHz e/ou 4 kHz foram submetidos à anamne se, audiometria tonal liminar, logoaudiometria e imitanciometria. A audiometria tonal e a logoaudiometria foram realizadas com os fones supra-aurais e com fones de inserção.

RESULTADOS:

Verificou-se que os limiares auditivos por via aérea obtidos com os fones de inserção foram menores em todas as frequências com exceção de 1 kHz. O intervalo aéreo-ósseo também foi menor com o fone de inserção, havendo um acoplamento aéreo-ósseo com este transdutor.

CONCLUSÃO:

Os fones de inserção ER-3A possibilitam melhora significativa nos limiares de audibilidade por via aérea em idosos, diminuindo ou até eliminando o intervalo aéreo-ósseo, sobretudo nas frequências altas, nos casos de colabamento do meato acústico externo.
ABSTRACT

PURPOSE:

To determine the efficacy of insert earphones in cases of collapsed external auditory canals.

METHODS:

Twenty-one elderly individuals with air-bone gaps greater than 10 dBHL at frequencies of 3 kHz and/or 4 kHz underwent anamnesis, pure tone audiometry, speech audiometry, and immittance test. Pure tone audiometry and speech audiometry were performed with both supra-aural and insert earphones.

RESULTS:

Air conduction thresholds were lower with insert earphones at all frequencies except 1 kHz. The magnitude of the air-bone gap was also lower with the insert earphones, and there was a air-bone coupling with this transducer.

CONCLUSION:

Insert earphones ER-3A allowed significant improvement in air conduction hearing thresholds in elderly individuals, reducing or even eliminating the air-bone gap in cases of collapsed external auditory canals, especially at higher frequencies.


Full text: Available Index: LILACS (Americas) Language: Portuguese Journal: Rev. Soc. Bras. Fonoaudiol Journal subject: Speech-Language Pathology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Language: Portuguese Journal: Rev. Soc. Bras. Fonoaudiol Journal subject: Speech-Language Pathology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de São Paulo/BR