Your browser doesn't support javascript.
loading
Controle de qualidade de monitores de diagnóstico por imagem e iluminância nos espaços de pós-processamento em serviços de imagiologia / Display quality control based on luminance measurements in post-processing units at imaging diagnosis centers
Pinto, Márcia; Pedro, Maria; Santos, António; Saraiva, António.
  • Pinto, Márcia; s.af
  • Pedro, Maria; s.af
  • Santos, António; Escola Superior de Tecnologia da Saúde de Coimbra. Coimbra. PT
  • Saraiva, António; Escola Superior de Tecnologia da Saúde de Coimbra. Coimbra. PT
Radiol. bras ; 45(1): 29-34, jan.-fev. 2012. ilus, tab
Article in Portuguese | LILACS | ID: lil-618392
RESUMO

OBJETIVO:

Avaliar a qualidade dos monitores de diagnóstico e tratamento de imagem digital em radiologia convencional (raios X) com base nos níveis de luminância. Fez-se a medição da iluminância dos postos de trabalho onde estavam os monitores.

MATERIAIS E MÉTODOS:

A medição da luminância foi realizada usando-se um detector Unfors Xi Light em cinco instituições, com 23 monitores primários e 22 secundários, sendo 6 usados em urgência, seguindo o método descrito no relatório TG18 da American Association of Physicists in Medicine (AAPM). O instrumento de medida utilizado foi o Delta Ohm HD 9221 luximeter.

RESULTADOS:

Para as recomendações da AAPM, tanto monitores primários como secundários cumpriram, no geral, os níveis de tolerância. No caso dos níveis de iluminância, nos locais de trabalho dos monitores primários encontravam-se ligeiramente acima do intervalo recomendado. Quanto às recomendações do The Royal College of Radiologists (RCR), os monitores que respeitaram os níveis recomendados estavam em minoria, embora as diferenças não fossem significativas.

CONCLUSÃO:

Em geral, os resultados foram satisfatórios, embora os monitores não fossem adequados para atingir as recomendações exigentes do RCR. Os níveis de referência entre os monitores primários e secundários deveriam aproximar-se, de modo a garantir a qualidade de imagem. Os monitores para visualização e diagnóstico de imagens médicas devem ser regularmente avaliados para manter um sistema com a qualidade exigida, dada a função que desempenham.
ABSTRACT

OBJECTIVE:

To evaluate the quality of primary and secondary displays of conventional radiography systems based on luminance measurements. The level of ambient lighting within the rooms was also measured. MATERIALS AND

METHODS:

Luminance measurements were performed with an Unfors Xi Light Detector in 23 primary and 22 secondary displays (six of them utilized in emergency departments) at five different institutions, and according to the method described on the American Association of Physicists in Medicine (AAPM) TG18 report. The level of ambient lighting was measured with Delta Ohm HD 9221 lux meter.

RESULTS:

In general, both the primary and secondary the displays were compliant with the AAPM guidelines. As regards ambient lighting, the primary workstation rooms were slightly above the recommended levels. As far as The Royal College of Radiologists (RCR) recommendations are concerned, the number of compliant displays corresponded to a minority, with no significant difference, though.

CONCLUSION:

Although most of the displays could not meet the strict RCR guidelines, in general, the results were satisfactory. The level of agreement between primary and secondary displays should be as high as possible in order to guarantee the image quality. Considering the key role played by diagnostic imaging systems, medical imaging displays must be regularly evaluated to assure that the quality of the imaging system is maintained.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Radiology / Technology, Radiologic / Diagnostic Equipment Type of study: Diagnostic study / Practice guideline Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Portugal Institution/Affiliation country: Escola Superior de Tecnologia da Saúde de Coimbra/PT

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Radiology / Technology, Radiologic / Diagnostic Equipment Type of study: Diagnostic study / Practice guideline Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Portugal Institution/Affiliation country: Escola Superior de Tecnologia da Saúde de Coimbra/PT