Fatores de correção volumétrica na medição de atividade de 99mTc e 123I em calibradores de radionuclídeos / Volume corrections factors in the measurement of 99mTc and 123I activities in radionuclide calibrators
Correia, Amanda Ribeiro; Iwahara, Akira; Tauhata, Luiz; Rezende, Eduarda Alexandre; Chaves, Tainá Olivieri; Oliveira, Antonio Eduardo de; Oliveira, Estela Maria de.
  • Correia, Amanda Ribeiro; Universidade Federal do Rio de Janeiro. Rio de Janeiro. BR
  • Iwahara, Akira; s.af
  • Tauhata, Luiz; s.af
  • Rezende, Eduarda Alexandre; Universidade Federal do Rio de Janeiro. Rio de Janeiro. BR
  • Chaves, Tainá Olivieri; s.af
  • Oliveira, Antonio Eduardo de; s.af
  • Oliveira, Estela Maria de; s.af
Radiol. bras ; 45(2): 93-97, mar.-abr. 2012. ilus, tab
Article in Portuguese | LILACS | ID: lil-624458
RESUMO

OBJETIVO:

Determinar os fatores de correção para a variação volumétrica dos radiofármacos contidos em recipientes de diferentes geometrias. Comparar a influência desses fatores na determinação da atividade de 99mTc e 123I utilizando dois tipos de calibradores um com câmara de ionização e outro com detector Geiger-Müller (G-M). Avaliar o desempenho de calibradores de alguns serviços de medicina nuclear em medição de atividade de 99mTc e 123I.

MATERIAIS E MÉTODOS:

Foram utilizados oito calibradores, frascos de vidro 10R, seringas de plástico de 3 e 5 mL e soluções de 99mTc e 123I. Os fatores de correção foram determinados a partir das medições práticas da variação da leitura do calibrador com a variação do volume da solução no recipiente. O desempenho foi avaliado em relação ao critério de aceitação de ±10% de exatidão exigida pela norma brasileira.

RESULTADOS:

A variação da resposta do calibrador com a variação do volume do frasco foi bem maior no calibrador que utiliza G-M. Ela também foi maior para 123I do que para 99mTc.

CONCLUSÃO:

Os resultados confirmam que a resposta dos calibradores depende do volume contido nos recipientes. Essa dependência é mais crítica para os calibradores equipados com detector G-M e para 123I quando comparado com 99mTc.
ABSTRACT

OBJECTIVE:

To determine correction factors for the variation in volume of radiopharmaceuticals in containers of different geometries, comparing the influence of such factors on the determination of 99mTc and 123I activity with two types of calibrators - one with ionization chamber and another with Geiger-Müller (G-M) detector -; and to evaluate calibrators performance in the measurement of 99mTc and 123I activities. MATERIALS AND

METHODS:

Eight calibrators, 10R glass vials, 3 and 5 mL plastic syringes and 99mTc and 123I solutions were utilized. The correction factors were determined with basis on practical measurements of the variation in the calibrators' response according to the volume of radionuclide solution in the glass vials. The performance was evaluated according to the acceptance criterion of ±10% accuracy required by the Brazilian standard.

RESULTS:

The variation of the calibrators' response according to the variation in radionuclide volume was reasonably greater in the calibrator with G-M detector. It was also greater for 123I than for 99mTc.

CONCLUSION:

The results confirm that the calibrators' response depends on the radionuclide volume contained in the vials. Such dependence is more critical for the calibrators equipped with G-M detector and for 123I as compared with 99mTc.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Radioisotopes / Technetium / Radiopharmaceuticals / Iofetamine / Nuclear Medicine Type of study: Practice guideline Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal do Rio de Janeiro/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Radioisotopes / Technetium / Radiopharmaceuticals / Iofetamine / Nuclear Medicine Type of study: Practice guideline Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal do Rio de Janeiro/BR