Your browser doesn't support javascript.
loading
Cierre de conducto arterioso persistente con dispositivo Nit-Occlud: Experiencia de 13 casos / Patent ductus arteriosus closure with a Nit-Occluded device: Experience with 13 cases
Munayer Calderón, Jaime E; Aldana Pérez, Tomás; Lázaro Castillo, José Luis; San Luis Miranda, Raúl; Ramírez Reyes, Homero; Amaya Hernández, Antonio; Carpio Hernández, Juan C.
  • Munayer Calderón, Jaime E; HG CMN La Raza. México. MX
  • Aldana Pérez, Tomás; HG CMN La Raza. México. MX
  • Lázaro Castillo, José Luis; HG CMN La Raza. México. MX
  • San Luis Miranda, Raúl; HG CMN La Raza. México. MX
  • Ramírez Reyes, Homero; HG CMN La Raza. México. MX
  • Amaya Hernández, Antonio; HG CMN La Raza. México. MX
  • Carpio Hernández, Juan C; HG CMN La Raza. México. MX
Arch. cardiol. Méx ; 75(4): 408-412, oct.-dic. 2005. ilus
Article in Spanish | LILACS | ID: lil-631919
RESUMEN
Se presenta la experiencia de 13 pacientes con diagnóstico de persistencia de conducto arterioso permeable tratados por vía percutánea con un dispositivo de nitinol (aleación de níquel y titanio), llamado Nit-Occlud. Fueron 10 mujeres y 3 hombres con edad promedio de 23.2 ±21.1 años. La angiografía en la aorta descendente previa al cierre reportó diámetro promedio del conducto arterioso de 3.8 ± 0.8 mm y la morfología del mismo fue de 10 conductos tipo A, 2 tipo E y 1 tipo C según la clasificación de Krichenko. El tamaño del oclusor seleccionado se basó en el diámetro de la ámpula aórtica y del sitio más estrecho del conducto. Se utilizaron 8 dispositivos 11x6 mm, 4 dispositivos 9x6 mm y 1 dispositivo 7x6 mm. Tres pacientes (25%) mostraron oclusión del 100% en forma inmediata (15 min después de la liberación del dispositivo), 7 (53%) presentaron fuga leve y difusa y sólo 2 pacientes (16.6%) mostraron fuga moderada a severa. Un paciente presentó migración del dispositivo a la rama izquierda de la arteria pulmonar (RIAP), se logró su rescate por vía percutánea enviándose el paciente posteriormente a cirugía. En el seguimiento ecocardiográfico a las 24 h 9 pacientes (69%) mostraron oclusión del 100%. A los 6 meses de seguimiento 10 tienen oclusión al 100% y sólo 2 pacientes tienen fuga residual leve (15%). Un paciente desarrolló cuadro de endocarditis infecciosa que finalmente provocó su deceso. Se concluye que el dispositivo Nit-Occlud es útil para el cierre de esta patología aunque hace falta mayor experiencia.
ABSTRACT
We report our experience with the percutaneous closure of patent ductus arteriosus with a Nit-Occlud device made of Nitinol (Nickel-Titanium alloy) in 13 patients; 10 women and 3 men, average age of 23.2 ±21.1 years. Average diameters were 3.8 + 0.8 mm and the morphologies, according to Krichenko's classification, were 10 type A, 2 type E, and 1 type C. The device was selected according to the aortic ampule and the narrowest part of the ductus. We implanted 8 occluders of 11 x 6 mm, 4 of 9 x 6 mm, and 1 of 7 x 6 mm. In three patients (25%) total occlusion was observed 15 minutes after implantation, in seven (54%) a trivial leak was observed, and in only two patients (16.6%) was the leak moderate to severe. In on patient, the occluder migrate to the pulmonary artery trunk and was successfully removed percutaneously; the patient was subjected then to surgical closure. Echocardiography follow-up 24 h later showed total occlusion in nine patients (69%). Six months after the procedure, two patient presented trivial leak (15%). One patient developed an endocarditic infection and died. We concluded that this occluder might be useful for this pathology, but further studies must be done.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Prostheses and Implants / Ductus Arteriosus, Patent Limits: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male Language: Spanish Journal: Arch. cardiol. Méx Journal subject: Cardiology Year: 2005 Type: Article Affiliation country: Mexico Institution/Affiliation country: HG CMN La Raza/MX

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Prostheses and Implants / Ductus Arteriosus, Patent Limits: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male Language: Spanish Journal: Arch. cardiol. Méx Journal subject: Cardiology Year: 2005 Type: Article Affiliation country: Mexico Institution/Affiliation country: HG CMN La Raza/MX