Your browser doesn't support javascript.
loading
Avaliação de exposições médicas em procedimentos pediátricos de radiologia intervencionista / Evaluation of medical radiation exposure in pediatric interventional radiology procedures
Navarro, Valéria Coêlho Costa; Navarro, Marcus Vinícius Teixeira; Maia, Ana Figueiredo; Oliveira, Adriano Dias Dourado; Oliveira, Aline da Silva Pacheco.
  • Navarro, Valéria Coêlho Costa; Universidade Federal de Sergipe. Salvador. BR
  • Navarro, Marcus Vinícius Teixeira; Instituto Federal de Educação, Ciência e Tecnologia da Bahia. Salvador. BR
  • Maia, Ana Figueiredo; Universidade Federal de Sergipe. Aracaju. BR
  • Oliveira, Adriano Dias Dourado; Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista. Salvador. BR
  • Oliveira, Aline da Silva Pacheco; Instituto Federal de Educação, Ciência e Tecnologia da Bahia. Salvador. BR
Radiol. bras ; 45(4): 210-214, jul.-ago. 2012. graf, tab
Article in Portuguese | LILACS | ID: lil-647861
RESUMO

OBJETIVO:

Avaliar as exposições pediátricas de radiologia intervencionista em dois hospitais do Estado da Bahia, visando contribuir para a construção de um cenário estadual e nacional, possibilitando o conhecimento das exposições e da necessidade de sua otimização, visto que as peculiaridades que envolvem a radiologia e a pediatria se potencializam quando se trata de procedimentos de radiologia intervencionista, em razão das doses elevadas de radiação.

MATERIAIS E MÉTODOS:

Foram avaliados 32 procedimentos em quatro salas nos dois principais hospitais que realizam procedimentos de radiologia intervencionista pediátrica na Bahia. Foram avaliados os valores de kerma no ar incidente e o produto kerma-área no ar de 27 procedimentos cardiológicos e 5 procedimentos cerebrais.

RESULTADOS:

Os valores máximos de produto kerma-área e kerma obtidos para procedimentos cardiológicos foram, respectivamente, 129,9 Gy.cm² e 947,0 mGy, e para procedimentos cerebrais, 83,3 Gy.cm² e 961,0 mGy.

CONCLUSÃO:

Os resultados deste estudo mostraram valores de exposições superiores em até 14 vezes os obtidos em estudos realizados em outros países, chegando próximos de resultados obtidos para procedimentos em adultos. Isto revela quão elevadas podem ser as exposições pediátricas, indicando a necessidade de constante otimização dos procedimentos e avaliação das exposições.
ABSTRACT

OBJECTIVE:

To evaluate pediatric radiation exposure in procedures of interventional radiology in two hospitals in the Bahia state, aiming at contributing to delineate the scenario at the state and national levels. The knowledge of exposure levels will allow an evaluation of the necessity of doses optimization, considering that peculiarities of radiology and pediatrics become even more significant in interventional radiology procedures which involve exposure to higher radiation doses. MATERIALS AND

METHODS:

A total of 32 procedures were evaluated in four rooms of the two main hospitals performing pediatric interventional radiology procedures in the Bahia state. Air kerma rate and kerma-area product were evaluated in 27 interventional cardiac and 5 interventional brain procedures.

RESULTS:

Maximum values for air kerma rate and kerma-area product and air kerma obtained in cardiac procedures were, respectively, 129.9 Gy.cm² and 947.0 mGy; and, for brain procedures were 83.3 Gy.cm² and 961.0 mGy.

CONCLUSION:

The present study results showed exposure values up to 14 times higher than those found in other foreign studies, and approximating those found for procedures in adults. Such results demonstrate excessive exposure to radiation, indicating the need for constant procedures optimization and evaluation of exposure rates.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Magnetic Resonance Imaging, Interventional / Cerebrum / Heart Limits: Adolescent / Child / Child, preschool / Female / Humans / Infant / Male / Infant, Newborn Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Instituto Federal de Educação, Ciência e Tecnologia da Bahia/BR / Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista/BR / Universidade Federal de Sergipe/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Magnetic Resonance Imaging, Interventional / Cerebrum / Heart Limits: Adolescent / Child / Child, preschool / Female / Humans / Infant / Male / Infant, Newborn Language: Portuguese Journal: Radiol. bras Journal subject: Diagnostic Imaging / Radiology Year: 2012 Type: Article Affiliation country: Brazil Institution/Affiliation country: Instituto Federal de Educação, Ciência e Tecnologia da Bahia/BR / Sociedade Brasileira de Hemodinâmica e Cardiologia Intervencionista/BR / Universidade Federal de Sergipe/BR