Your browser doesn't support javascript.
loading
Videoenteroscopia de balón único con prototipo Olympus: experiencia preliminar / Single-balloon videoenteroscopy olympus prototype: preliminary experience
Landaeta, J; Rodríguez, M; Días, C; Urdaneta, C; Casanova, R.
  • Landaeta, J; s.af
  • Rodríguez, M; s.af
  • Días, C; s.af
  • Urdaneta, C; s.af
  • Casanova, R; s.af
GEN ; 65(3): 204-206, sep. 2011. tab
Article in Spanish | LILACS | ID: lil-664148
RESUMEN

Introducción:

La enteroscopia de balón único es una técnica segura con alto alcance diagnóstico y terapéutico. Sin embargo, existe la dificultad para el manejo de accesorios mientras mayor es la profundidad de inserción.

Objetivo:

Mostrar ventajas en el diagnóstico y tratamiento de patología del intestino delgado utilizando un prototipo Olympus con canal 3,2 mm. Pacientes y

métodos:

18 procedimientos en 13 pacientes. Diciembre 2009-Abril 2010. Se utilizó el prototipo Olympus X SIF 180 JY, con canal 3.2 mm y canal externo de irrigación. Se usaron accesorios endoscópicos de tecnología convencional para colonoscopia. Se registraron datos demográficos, indicaciones, vía de acceso, tiempo del procedimiento, profundidad de inserción, hallazgos y terapéutica.

Resultados:

13 pacientes 9 femenino, 4 masculino, edad promedio 47,61 años. Indicaciones principales sangrado digestivo oscuro (4), Diarrea crónica (4), síndrome de poliposis (3). Procedimientos. Vía anterógrada 13, retrógrada 4, combinado 1. Tiempo promedio 40,46 minutos anterógrada, 36,35 minutos retrógrada. Profundidad de inserción 201,53 cm anterógrada, 95 cm retrograda. Estudio normal 46,15 % de los pacientes, hallazgos positivos 53,84%. Terapéutica 5 (27,7%) clips (2), inyección con adrenalina+argón (1), polipectomía (1), dilatación de estenosis (1). No hubo complicaciones.

Conclusión:

el prototipo Olympus X SIF 180 JY cuenta con canal terapéutico que permite la utilización de accesorios endoscópicos de tecnología convencional con menor dificultad de deslizamiento a mayor profundidad de inserción y canal de irrigación externo que mejora la visibilidad del campo. Ambas características favorecerían la realización de procedimientos terapéuticos.
ABSTRACT

Introduction:

The single balloon enteroscopy is a safe technique with high diagnostic and therapeutic yield. However, there is the difficulty to manage attachments while the greater the depth of insertion.

Objective:

To show benefits in the diagnosis and treatment of diseases of the small bowel using a prototype Olympus with 3.2 mm channel. Patients and

Methods:

18 procedures in 13 patients. December 2009-April 2010. We used the prototype Olympus SIF 180 JY X, with 3.2 mm channel and external canal irrigation. Standard endoscopic accessories were used for colonoscopy. Recorded demographic data, indications, route of access, procedure time, insertion depth, and therapeutic findings.

Results:

13 patients 9 women, 4 men, mean age 47.61 years. Main indications obscure gastrointestinal bleeding (4), chronic diarrhea (4), polyposis syndrome (3). Procedures. 13 antegrade, 4 retrograde, combined 1. Average Time 40.46 minutes antegrade, retrograde 36.35 minutes. Depth of insertion 201.53 cm antegrade, retrograde 95 cm. Study normal 46.15% of patients, positive

findings:

53.84%, with active bleeding lymphangioma (1), Dieulafoy lesion (1), celiac disease (1), Crohn's (1), polyposis (3). Therapeutics 5 (27.7%) clips (2), adrenaline injection + argon (1), polypectomy (1), dilatation of stenosis (1). There were no complications.

Conclusion:

The prototype Olympus SIF 180 JY X has therapeutic channel allows the use of standard endoscopic accessories more easily slip deeper insertion and external irrigation canal which enhances the visibility of the field. Both features favor the therapeutic procedures.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Incidental Findings / Double-Balloon Enteroscopy / Intestine, Small Type of study: Diagnostic study / Evaluation studies Limits: Adult / Female / Humans / Male Language: Spanish Journal: GEN Year: 2011 Type: Article

Similar

MEDLINE

...
LILACS

LIS