Comparison of methods to evaluate implant-abutment interface
Faria, Karina Oliveira; Silveira-Júnior, Clébio Domingues; Silva-Neto, João Paulo; Mattos, Maria da Glória Chiarello; Silva, Marlete Ribeiro; Universidade Federal de Uberlândia.
  • Faria, Karina Oliveira; Universidade Federal de Uberlândia. Faculdade de Odontologia. Uberlândia. BR
  • Silveira-Júnior, Clébio Domingues; Universidade Federal de Uberlândia. Faculdade de Odontologia. Uberlândia. BR
  • Silva-Neto, João Paulo; Universidade de Campinas. Faculdade de Odontologia. Piracicaba. BR
  • Mattos, Maria da Glória Chiarello; Universidade de São Paulo. Faculdade de Odontologia. Ribeirão Preto. BR
  • Silva, Marlete Ribeiro; Universidade de Campinas. Faculdade de Odontologia. Uberlândia. BR
Braz. j. oral sci ; 12(1): 37-40, jan.-mar. 2013. ilus, tab
Article in English | LILACS, BBO | ID: lil-671930
ABSTRACT

Aim:

To compare two main methods of two-dimensional measurement of fit at the implant prosthodontic interface, testing the hypothesis that optical microscopy (OM) can reliably and efficiently scanning electronic microscopy (SEM).

Methods:

Four frameworks with four titanium abutments joined with titanium bars were used. The implant-abutment interfaces were examined by three different methods, forming 3 groups analysis by OM (40x), and analysis by SEM at 300x and 500x. Readings were taken at the mesial and distal proximal surfaces on the horizontal and vertical axes of each implant (n=32). One-way ANOVA with a significance level of 5% was used for statistical analysis.

Results:

Neither the horizontal fit nor vertical fit values of the 3 groups presented statistically significant differences (p=0.410 and p=0.543).

Conclusions:

OM was found to be an accurate two-dimensional method for abutment-framework or implant-abutment interface measurements, with lower costs than SEM. SEM micrographs at 500x presented technical difficulties for the readings that might produce different results.
Subject(s)
Full text: Available Index: LILACS (Americas) Main subject: Microscopy, Electron, Scanning / Dental Implants / Methodology as a Subject Language: English Journal: Braz. j. oral sci Journal subject: Dentistry Year: 2013 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de Uberlândia/BR / Universidade de Campinas/BR / Universidade de São Paulo/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Microscopy, Electron, Scanning / Dental Implants / Methodology as a Subject Language: English Journal: Braz. j. oral sci Journal subject: Dentistry Year: 2013 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade Federal de Uberlândia/BR / Universidade de Campinas/BR / Universidade de São Paulo/BR