Caracterização micrométrica das superfícies dos implantes das cinco maiores companhias do mercado brasileiro. Parte III: implantes SIN / Characterization of implants surface of the five largest companies in the Brazilian market, on micrometric level. Part III: SIN implants
Rosa, Márcio Borges; Albrektsson, Tomas; Francischone, Carlos Eduardo; Schwartz Filho, Humberto Osvaldo; Wennerberg, Ann.
  • Rosa, Márcio Borges; São Leopoldo Mandic. Campinas. BR
  • Albrektsson, Tomas; Universidade de Gotemburgo. Gotemburgo. SE
  • Francischone, Carlos Eduardo; USP. FOB. São Paulo. BR
  • Schwartz Filho, Humberto Osvaldo; Unesp. São Paulo. BR
  • Wennerberg, Ann; Universidade de Malmo. Malmo. BR
Dent. press implantol ; 6(3): 60-70, jul.-set. 2012. ilus, graf
Article in Portuguese | LILACS, BBO | ID: lil-681720
RESUMO

Introdução:

a qualidade da interface osso-implante é influenciada diretamente pela rugosidade da superfície do implante; e uma rugosidade média, com Sa entre 1 e 2µm, tem demonstrado melhores resultados clínicos e laboratoriais. No Brasil, são instalados mais de dois milhões de implantes por ano, sendo que 79% são fabricados por empresas nacionais. Porém, muito pouco é divulgado ou se conhece sobre a caracterização das superfícies desses implantes, a nível micrométrico. Esse estudo visa avaliar e caracterizar, numericamente, a superfície dos implantes da SIN (Sistema de Implante Nacional), uma das cinco maiores empresas do mercado brasileiro.

Métodos:

foram avaliados 6 implantes, comprados diretamente no mercado, de 2 desenhos da companhia (Tryon-HE e Strong-SW) e de diferentes lotes, através de um interferômetro de luz. Foram realizados 9 medições, escolhidas aleatoriamente, para cada unidade, sendo 3 nos topos, 3 nos vales e 3 nos flancos das roscas. O mesmo padrão foi seguido para avaliação através microscópio eletrônico de varredura.

Resultados:

os implantes analisados dessa companhia, apresentaram valores de Sa de 0,84µm para o Tryon-HE e de 1,01µm para o Strong SW. Na comparação entre os lotes, somente o desenho SW apresentou diferença estatisticamente significativa entre si.

Conclusões:

os valores de rugosidade encontrados classificam as superfície, dos implantes Tryon-HE como minimamente rugosas e dos implantes Strong-SW como moderadamente rugosas.
ABSTRACT

Introduction:

The quality of the bone-implant interface is directly influenced by implant surface roughness and a roughness average, with the Sa between 1 to 2µm, has demonstrated better clinical and laboratory results. In Brazil, more than two million implants per year are installed, where 79% are manufactured by domestic companies. However, very little is known or published about the characterization of surfaces of these implants, on the micrometer level. The aims of this study are to evaluate and characterize numerically the surface of the implants of SIN (Sistema de Implante Nacional) company, one of the five largest companies in the Brazilian market.

Methods:

Were evaluated a total of 6 implants, purchased directly on the market, of two different designs (Tryon-HE and Strong-SW) and different batches, using light interferometry. Were performed 9 measurements randomly chosen for each unit,3 on the tops, 3 on the valleys and 3 on the flanks of the threads. The same pattern was followed for evaluation by scanning electron microscope.

Results:

the analyzed implants from this company showed Sa values of 0.84µm for Tryon-HE and 1.01µm for Strong SW. Comparing the batches, only the SW design showed statistically significant differences between them.

Conclusions:

the roughness values found herein categorize the surfaces of Tryon-HE as minimally rough, and Strong-SW implants as moderately rough.
Subject(s)

Search on Google
Index: LILACS (Americas) Main subject: Materials Testing / Dental Implants Type of study: Controlled clinical trial Country/Region as subject: South America / Brazil Language: Portuguese Journal: Dent. press implantol Journal subject: Dentistry Year: 2012 Type: Article Affiliation country: Brazil / Sweden Institution/Affiliation country: São Leopoldo Mandic/BR / USP/BR / Unesp/BR / Universidade de Gotemburgo/SE / Universidade de Malmo/BR

Similar

MEDLINE

...
LILACS

LIS

Search on Google
Index: LILACS (Americas) Main subject: Materials Testing / Dental Implants Type of study: Controlled clinical trial Country/Region as subject: South America / Brazil Language: Portuguese Journal: Dent. press implantol Journal subject: Dentistry Year: 2012 Type: Article Affiliation country: Brazil / Sweden Institution/Affiliation country: São Leopoldo Mandic/BR / USP/BR / Unesp/BR / Universidade de Gotemburgo/SE / Universidade de Malmo/BR