RESUMO
FUNDAMENTO Apesar da evolução tecnológica do cardiodesfibrilador implantável, uma das questões que permanece em aberto é sobre o eventual benefício do cardiodesfibrilador implantável de dupla câmara comparativamente ao de câmara única, para diminuir os choques inapropriados.

OBJETIVO:

Avaliar, em pacientes com cardiodesfibrilador implantável, qual é o tipo de dispositivo que proporciona menor número de choques inapropriados (câmara dupla versus câmara única).

MÉTODOS:

Meta-análise da literatura publicada, de estudos randomizados, comparando cardiodesfibrilador implantável de câmara dupla ao de câmara única, que tenham, como endpoint avaliado, a ocorrência de choques inapropriados.

RESULTADOS:

O cardiodesfibrilador implantável de câmara dupla não mostrou benefícios na redução do número de pacientes com choques inapropriados. Pelo contrário, na análise de efeitos fixos, a associação foi tendencialmente favorável ao cardiodesfibrilador implantável de câmara única (OR = 1,53; IC95% 0,91-2,57), não obstante a ausência de significado estatístico (p = 0,11). Merece destaque a heterogenia observada nos resultados (I2 = 53%), o que motivou a replicação da análise utilizando um modelo de efeitos aleatórios. No entanto, as diferenças significativas permaneceram na ocorrência de choques inapropriados em ambos os grupos (OR = 1,1; IC95% 0,37-3,31; p = 0,86). Para complementar a análise, procedeu-se à análise de sensibilidade, na qual se verificou que a exclusão de um estudo resultava na mais baixa heterogenia observada (I2=24%) e na associação com os choques inapropriados significativamente favorável ao cardiodesfibrilador de câmara única (OR = 1,91; IC95% 1,09-3,37; p = 0,27).

CONCLUSÕES:

Verificou-se a não existência de evidência clara de superioridade de qualquer um dos dispositivos avaliados.
ABSTRACT

BACKGROUND:

Despite the technological evolution of the implantable defibrillator, one of the questions that remains is the possible benefit of the dual chamber versus single chamber implantable cardioverter defibrillator (ICD) in reducing inappropriate shocks.

OBJECTIVE:

To evaluate which type of device provides fewer inappropriate shocks (dual chamber versus single chamber) in patients with implantable cardioverter defibrillators (ICDs).

METHODS:

Meta-analysis of randomized studies published in the literature comparing dual-chamber implantable cardioverter defibrillators to single chamber devices which have been known to cause, as an evaluated endpoint, inappropriate shocks.

RESULTS:

The dual-chamber implantable cardioverter showed no benefit in reducing the number of inappropriate shocks. In fact, the opposite was shown. In the analysis of fixed effects, the association tended to favor single-chamber implantable cardioverter defibrillators (OR = 1.53, CI 95% 0.91-2.57), despite the absence of statistical significance (p = 0.11). We highlight the heterogeneity observed in the results (I² = 53%), which motivated a replication of the analysis using a model of random effects. However, significant differences remained in the occurrence of inappropriate shocks in both groups (OR = 1.1, 95% CI 0.37-3.31; p = 0.86). To complement the analysis, we proceeded to perform sensitivity analysis, which showed that the exclusion of a study resulted in the lowest heterogeneity observed (I²=24%) and the association with inappropriate shocks significantly favored the single chamber cardiodefibrillator (OR = 1.91; 95% CI 1.09-3.37; p = 0.27).

CONCLUSIONS:

It was determined that there was no clear evidence of superiority of any of the devices evaluated.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Electric Countershock / Defibrillators, Implantable Type of study: Controlled clinical trial / Etiology study / Risk factors / Systematic reviews Limits: Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2013 Type: Article Affiliation country: Portugal

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Electric Countershock / Defibrillators, Implantable Type of study: Controlled clinical trial / Etiology study / Risk factors / Systematic reviews Limits: Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2013 Type: Article Affiliation country: Portugal