Há risco de êmbolos durante a infusão com o uso de aquecedores com linhas de aquecimento de sangue e líquidos? / Is there risk of emboli during Infusion with Line type blood-liquid warmers? / ¿Existen riesgos de émbolos durante la infusión con el uso de calentadores con líneas de calentamiento de sangre y líquidos?
Erkin, Yüksel; Taşdöğen, Aydın; Gönüllü, Edip.
  • Erkin, Yüksel; Dokuz Eylül University. Faculdade de Medicina. Departamento de Anestesiologia e Reanimação. Izmir. TR
  • Taşdöğen, Aydın; Dokuz Eylül University. Faculdade de Medicina. Departamento de Anestesiologia e Reanimação. Izmir. TR
  • Gönüllü, Edip; Dokuz Eylül University. Faculdade de Medicina. Departamento de Anestesiologia e Reanimação. Izmir. TR
Rev. bras. anestesiol ; 63(5): 389-392, set.-out. 2013. ilus, tab
Article in Portuguese | LILACS | ID: lil-691371
RESUMO
JUSTIFICATIVA E

OBJETIVOS:

Os aquecedores de sangue e de líquidos que aquecem em linha são amplamente usados por causa do baixo custo, do uso prático e porque não dependem de equipamentos. Nosso objetivo foi investigar a formação de bolhas em dois aquecedores tipo linha com duas formas de aquecimento diferentes.

MATERIAIS E MÉTODOS:

Dois grupos foram designados às marcas de aquecedores de sangue e líquidos S-line e Astoflo®. Com o uso de 10 conjuntos de soro para cada grupo (n = 20), 1.000 mL de solução NaCl a 9% foram infundidos a 350 mL.hora-1 durante uma hora na sala de operação. As seguintes temperaturas foram mensuradas das partes proximal, intermediária e distal das linhas; do ambiente de ensaio; do líquido usado e do líquido ao atingir a cânula após o aquecimento. O tempo para a formação visível de bolhas foi registrado. Os achados foram estatisticamente comparados com o uso do teste-U de Mann-Whitney.

RESULTADOS:

Não houve diferença entre os grupos em relação às temperaturas proximal, intermediária e distal das partes das linhas; do ambiente do estudo; do líquido usado e do líquido ao atingir a cânula (p > 0,05). Bolhas foram observadas nos dois aquecedores e o tempo para a formação de bolhas foi semelhante nos dois grupos de estudo (p = 0,143).

CONCLUSÕES:

No cenário experimental, criamos condições semelhantes ao nosso ambiente clínico. Ambos os tipos de aquecedores forneceram níveis de aquecimento semelhantes e formaram bolhas visíveis. Considerando que uma pequena quantidade de êmbolos pode ser fatal em bebês e crianças, a formação de bolhas deve ser seriamente considerada em caso de êmbolos e estudos adicionais devem ser feitos para determinar a quantidade, as razões e os conteúdos da formação de bolhas.
ABSTRACT

INTRODUCTION:

Line type blood-liquid warmers are used widely due to their low expense, practical use and nondependence on sets. We aimed to investigate the relationship of bubbles in line type warmers with two different warming properties. MATERIALS AND

METHODS:

Two groups were designed with S-line and Astoflo® brand blood-liquid warmers. By using 10 medisets for each group (n = 20), we infused 1,000 mL 0.9% NaCl solutions at 350 mL.hour-1 speed for one hour in the operating room. Temperatures at the proximal, midway and distal parts of lines, temperature of experiment environment, temperature of liquid used and temperature of liquid reaching the cannula after warming were measured. Time to visually observable bubble formation was recorded. We compared findings statistically using the Mann-Whitney U test.

RESULTS:

There were no differences between the groups with respect to temperatures at the proximal, midway and distal parts of lines, temperature of experiment environment, temperature of liquid used and temperature of liquid reaching the cannula (p > 0.05). Bubbles were observed with both warmers and time to bubble formation was similar in the two study groups (p = 0.143).

CONCLUSIONS:

In the experimental setting, we have designed conditions similar to our clinical environment. Both types of warmers provided similar warming levels and formed visible bubbles. Considering that low amounts of emboli can be fatal in infants and children, bubble formation should be taken seriously into account for emboli and further studies should be carried out to determine the amount, the reasons and the contents of bubble formation.
RESUMEN
JUSTIFICATIVA Y

OBJETIVOS:

Los calentadores que poseen un calentamiento en línea de sangre y de líquidos son extensamente usados a causa de su bajo coste, del uso práctico y también porque no dependen de equipos. Nuestro objetivo fue investigar la formación de burbujas en dos calentadores de tipo línea con dos formas de calentamiento diferentes. MATERIALES Y MÉTODOS Dos grupos fueron designados a las marcas de calentadores de sangre y líquidos S-line y Astoflo®. Con el uso de 10 conjuntos de suero para cada grupo (n = 20), 1.000 mL de solución NaCl al 9% se infundieron a 350 mL.hora-1 durante una hora en quirófano. Las siguientes temperaturas fueron mensuradas de las partes proximal, intermedia y distal de las líneas; del ambiente de ensayo; del líquido usado y del líquido al alcanzar la bránula después del calentamiento. El tiempo para la formación visible de burbujas fue registrado. Los hallazgos fueron estadísticamente comparados con el uso del test-U de Mann-Whitney.

RESULTADOS:

No hubo diferencias entre los grupos con relación a las temperaturas proximal, intermedia y distal de las partes de las líneas; del ambiente del estudio; del líquido usado y del líquido al alcanzar la bránula (p > 0,05). Burbujas se observaron en los dos calentadores y el tiempo para la formación de burbujas fue similar en los dos grupos de estudio (p = 0,143).

CONCLUSIONES:

Dentro del escenario experimental, creamos condiciones parecidas a nuestro ambiente clínico. Ambos tipos de calentadores suministraron niveles de calentamiento parecidos y formaron burbujas visibles. Considerando que una pequeña cantidad de émbolos puede ser fatal en bebés y en niños, la formación de burbujas debe ser seriamente considerada en caso de émbolos y estudios adicionales deben ser realizados para determinar la cantidad, las razones y los contenidos de la formación de burbujas.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Rewarming / Embolism Type of study: Etiology study / Risk factors Limits: Humans Language: Portuguese Journal: Rev. bras. anestesiol Journal subject: Anesthesiology Year: 2013 Type: Article Affiliation country: Turkey Institution/Affiliation country: Dokuz Eylül University/TR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Rewarming / Embolism Type of study: Etiology study / Risk factors Limits: Humans Language: Portuguese Journal: Rev. bras. anestesiol Journal subject: Anesthesiology Year: 2013 Type: Article Affiliation country: Turkey Institution/Affiliation country: Dokuz Eylül University/TR