Your browser doesn't support javascript.
loading
Análise de corpos de prova mandibulares utilizados para avaliação de sistema de fixação óssea / Analysis of mandibular test specimens used to assess a bone fixation system
Campinas; s.n; jan. 2013. 52 p. ilus, tab, graf.
Thesis in Portuguese | LILACS | ID: lil-691887
RESUMO

Objetivo:

O objetivo deste trabalho é avaliar, por meio de testes biomecânicos, se materiais sintéticos diferentes, empregados para a confecção de corpos de prova apresentam comportamento biomecânico diferente, em comparação aos demais, ao simular in vitro a resistência ao carregamento de um método de fixação consagrado para OSRM.

Material:

30 réplicas de hemimandíbulas humana, sintéticas e padronizadas, com OSRM, divididas em três grupos de 10 amostras cada Grupo A - plástico ABS, Grupo B - Poliamida, e Grupo C - Poliuretano. Estas foram fixadas por três parafusos bicorticais posicionais (16 mm de comprimento, sistema 2.0 mm), em disposição "L" invertido, utilizando-se guias de perfuração e avanço de 5 mm.

Métodos:

Cada amostra foi submetida ao carregamento linear vertical e os valores de resistência ao carregamento registrado em 1, 3, 5, 7 e 10 mm de deslocamento. As médias e desvio padrão foram comparados, empregando-se análise de variância (p<0.05) e pelo teste Tukey.

Resultados:

Notou-se uma tendência de menores valores no grupo B em relação aos grupos A e C. Nos deslocamentos 3 e 5 mm, houve uma diferença entre os grupos A e C para o grupo B (p<0.05). Nos deslocamentos 7 e 10 mm houve diferença entre os 3 grupos sendo os maiores valores encontrados no grupo C e os menores no grupo B (p<0.05).

Conclusões:

Levando-se em consideração os resultados obtidos e o comportamento de cada material utilizado como substrato, podemos considerar que ocorreram diferenças significativas entre os materiais, quando comparados entre si.
ABSTRACT

Purpose:

The aim of this study was to assess through biomechanical testing if different synthetic materials used to fabricate test specimens have a different biomechanical behavior in comparison with other materials when simulating in vitro load resistance of a fixation method established for SSRO.

Material:

Thirty synthetic and standardized human hemimandible replicas with SSRO were divided into three groups of 10 samples each. Group A - ABS plastic; Group B - Polyamide and Group C - Polyurethane. These were fixated with three bicortical position screws (16 mm in length, 2.0mm system) in an inverted L pattern using perforation guide and 5 mm advancement.

Methods:

Each sample was submitted to linear vertical load and load strength values were recorded at 1, 3, 5, 7 and 10 mm of displacement. The means and standard deviation were compared using the Analysis of Variance (p<0.05) and the Tukey's test.

Results:

A tendency for lower values was observed in Group B in comparison with Groups A and C. At 3 and 5 mm of displacement, a difference between Groups A and C was found in comparison with Group B (p<0.05). At 7 and 10 mm of displacement a difference was found among the 3 groups, in which Group C showed the highest values and Group B the lowest (p<0.05).

Conclusions:

Taking into consideration the results obtained and the behavior of each material used as a substrate, significant differences occurred among the materials when compared among them.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Osteotomy, Sagittal Split Ramus / Fracture Fixation Limits: Humans Language: Portuguese Year: 2013 Type: Thesis

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Osteotomy, Sagittal Split Ramus / Fracture Fixation Limits: Humans Language: Portuguese Year: 2013 Type: Thesis