Angiografía no contrastada con Arterial Spin Labeling / Non-contrast angioresonance with Arterial Spin Labeling
Delgado, Jorge; Rascovsky, Simón; Calvo, Victor; López, Fredy; Castrillón, Gabriel; Foerster, Bernd.
  • Delgado, Jorge; Instituto de Alta Tecnología Médica. Medellín. CO
  • Rascovsky, Simón; Instituto de Alta Tecnología Médica. Medellín. CO
  • Calvo, Victor; Instituto de Alta Tecnología Médica. Medellín. CO
  • López, Fredy; Instituto de Alta Tecnología Médica. Medellín. CO
  • Castrillón, Gabriel; Instituto de Alta Tecnología Médica. Medellín. CO
  • Foerster, Bernd; Philips Medical Systems. BR
Rev. chil. radiol ; 19(3): 109-113, 2013. ilus
Article in Spanish | LILACS | ID: lil-695018
RESUMEN

Introducción:

La angiografía por resonancia magnética no contrastada realizada con "Arterial Spin Labeling" (ARM ASL) es un método diseñado para marcar los espines sanguíneos y así crear un contraste endógeno adecuado para evaluar territorios vasculares selectivamente sin la necesidad de aplicar medio de contraste intravenoso (compuestos de Gadolinio).

Metodología:

Se realizó un estudio descriptivo de una serie de casos, donde se describen los detalles técnicos y los resultados de la aplicación de la ARM ASL en equipos de 1.5 y 3 Tesla en voluntarios sanos.

Resultados:

Se observaron dos casos Para la técnica angiográfica del primer caso (ASL "Flow-in") se usó un resonador de 3T, sincronización cardiaca, una secuencia b-SSFP 3D y un pre-pulso de inversión, este último para saturar los tejidos estáticos. El volumen de examen se ubicó en el plano axial teniendo la precaución de cubrir la anatomía vascular renal, lo cual se logra en la mayoría de los casos con 60 a 70 cortes de 2 mm solapados en 50 porciento, voxel de 2x1x1 mm y campo de visión (FOV) de 250x100 mm. El protocolo del segundo caso fue obtenido en un equipo de 1.5T, sin sincronización cardiaca, con un navegador respiratorio dia fragmático y con una secuencia coronal Turbo SE 3D después de aplicar dos pre-pulsos de marcación sanguínea, el primero similar al del caso anterior y el segundo, o pulso selectivo, para marcar el flujo del vaso de interés. Con este método (ASL "Flow-Out") sólo la sangre marcada emite señal.

Conclusión:

Las técnicas de angiografía b-SSFP 3D y Turbo SE 3D no contrastadas con pre-pulsos de ASL en 1.5 y 3T son alternativas disponibles y, por lo tanto, pueden considerarse como complemento a otros métodos de angiografía por resonancia magnética al momento de evaluar la patología vascular.
ABSTRACT

Introduction:

Non-contrast magnetic resonance angiography using "Arterial Spin Labeling" (MRA ASL) is a technique designed to label blood spins and therefore create an endogenous contrast suitable for selectively evaluating vascular territories without intravenous contrast (Gadolinium compounds).

Methodology:

Technical details and results of the implementación of the MRA ASL using 1.5 and 3.0 Tesla systems in healthy volunteers is described.

Results:

Two cases were observed for the angiographic technique of the first case (ASL "Flow-in") in a 3.0 T unit, cardiac synchronization (cardiac gating), a 3D b-SSFP sequence, and an inversion pre-pulse was used, the latter to saturate the static tissues. The examination volume was located in the axial plane taking care to cover the renal vascular anatomy, which is achieved in most cases with 60 to 70 2 mm slices overlapped in 50%, voxel of 2x1x1 mm and a field of vision (FOV) of 250 x100 mm. The protocol for the second case was obtained on a 1.5 T system, without cardiac gating, with a diaphragmatic respiratory navigator and a 3D Turbo SE coronal sequence after applying two pre-pulse blood saturation bands, the first similar to the previous case and the second, or selective pulse, to label the flow of the vessel of interest. With this method (ASL "Flow-Out") only the labeled blood emits a signal.

Discussion:

3D b-SSFP and 3D Turbo SE non-contrast angiography techniques with ASL pre-pulses in 1.5 and 3T are available alternatives and, therefore, can be considered as a complement to other methods of magnetic resonance angiography when assessing vascular pathology.
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Spin Labels / Magnetic Resonance Angiography Limits: Humans Language: Spanish Journal: Rev. chil. radiol Journal subject: Radiology Year: 2013 Type: Article Affiliation country: Brazil / Colombia Institution/Affiliation country: Instituto de Alta Tecnología Médica/CO / Philips Medical Systems/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Spin Labels / Magnetic Resonance Angiography Limits: Humans Language: Spanish Journal: Rev. chil. radiol Journal subject: Radiology Year: 2013 Type: Article Affiliation country: Brazil / Colombia Institution/Affiliation country: Instituto de Alta Tecnología Médica/CO / Philips Medical Systems/BR