Espessura do tecido subcutaneo como preditor independente de ruido em imagem de TC cardiaca / Subcutaneous Tissue Thickness is an Independent Predictor of Image Noise in Cardiac CT
Staniak, Henrique Lane; Sharovsky, Rodolfo; Pereira, Alexandre Costa; Castro, Claudio Campi de; Bensenor, Isabela M.; Lotufo, Paulo A.; Bittencourt, Marcio Sommer.
  • Staniak, Henrique Lane; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Sharovsky, Rodolfo; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Pereira, Alexandre Costa; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Castro, Claudio Campi de; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Bensenor, Isabela M.; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Lotufo, Paulo A.; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
  • Bittencourt, Marcio Sommer; Universidade de Sao Paulo. Hospital Universitario. Sao Paulo. BR
Arq. bras. cardiol ; 102(1): 86-92, 1/2014. tab, graf
Article in Portuguese | LILACS | ID: lil-704041
RESUMO
Fundamento Há poucos dados sobre a definição de parâmetros simples e robustos para predizer artefato de imagem em tomografia computadorizada (TC) cardíaca.

Objetivos:

Avaliar o valor da simples medida da espessura do tecido subcutâneo (espessura pele-esterno) como preditor de artefato de imagem em TC cardíaca.

Métodos:

O estudo avaliou 86 pacientes submetidos a angiotomografia computadorizada cardíaca (ATCC) com sincronização prospectiva com ECG e avaliação de escore de cálcio coronário com 120 kV e 150 mA. A qualidade da imagem foi medida objetivamente pelo artefato de imagem na aorta em ATCC, sendo 'artefato baixo' definido como aquele < 30 UH. Os diâmetros torácicos anteroposterior e laterolateral, o artefato de imagem na aorta e a espessura pele-esterno foram medidos como preditores de artefato em ATCC. A associação de preditores e artefato de imagem foi avaliada usando-se correlação de Pearson.

Resultados:

A dose média de radiação foi 3,5 ± 1,5 mSv. O artefato de imagem médio na ATCC foi de 36,3 ± 8,5 UH, sendo o artefato de imagem médiona fase sem contraste do exame de 17,7 ± 4,4 UH. Todos os preditores foram independentemente associados com artefato em ATCC. Os melhores preditores foram espessura pele-esterno, com correlação de 0,70 (p < 0,001), e artefato de imagem na fases em contraste,com correlação de 0,73 (p < 0,001). Ao avaliar a habilidade de predizer artefato de imagem baixo, as áreas sob a curva ROC para o artefato de imagem na fases em contraste e para a espessura pele-esterno foram 0,837e 0,864, respectivamente.

Conclusão:

Tanto espessura pele-esterno quanto artefato de escore de cálcio são preditores simples e precisos de artefato de imagem em ATCC. Tais parâmetros podem ser incorporados aos protocolos ...
ABSTRACT

Background:

Few data on the definition of simple robust parameters to predict image noise in cardiac computed tomography (CT) exist.

Objectives:

To evaluate the value of a simple measure of subcutaneous tissue as a predictor of image noise in cardiac CT.

Methods:

86 patients underwent prospective ECG-gated coronary computed tomographic angiography (CTA) and coronary calcium scoring (CAC) with 120 kV and 150 mA. The image quality was objectively measured by the image noise in the aorta in the cardiac CTA, and low noise was defined as noise < 30HU. The chest anteroposterior diameter and lateral width, the image noise in the aorta and the skin-sternum (SS) thickness were measured as predictors of cardiac CTA noise. The association of the predictors and image noise was performed by using Pearson correlation.

Results:

The mean radiation dose was 3.5 ± 1.5 mSv. The mean image noise in CT was 36.3 ± 8.5 HU, and the mean image noise in non-contrast scan was 17.7 ± 4.4 HU. All predictors were independently associated with cardiac CTA noise. The best predictors were SS thickness, with a correlation of 0.70 (p < 0.001), and noise in the non-contrast images, with a correlation of 0.73 (p < 0.001). When evaluating the ability to predict low image noise, the areas under the ROC curve for the non-contrast noise and for the SS thickness were 0.837 and 0.864, respectively.

Conclusion:

Both SS thickness and CAC noise are simple accurate predictors of cardiac CTA image noise. Those parameters can be incorporated in standard CT protocols to adequately adjust radiation exposure. .
Subject(s)


Full text: Available Index: LILACS (Americas) Main subject: Aortography / Tomography, X-Ray Computed / Coronary Angiography / Artifacts / Subcutaneous Tissue Type of study: Diagnostic study / Evaluation studies / Prognostic study / Risk factors Limits: Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2014 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade de Sao Paulo/BR

Similar

MEDLINE

...
LILACS

LIS


Full text: Available Index: LILACS (Americas) Main subject: Aortography / Tomography, X-Ray Computed / Coronary Angiography / Artifacts / Subcutaneous Tissue Type of study: Diagnostic study / Evaluation studies / Prognostic study / Risk factors Limits: Aged / Female / Humans / Male Language: Portuguese Journal: Arq. bras. cardiol Journal subject: Cardiology Year: 2014 Type: Article Affiliation country: Brazil Institution/Affiliation country: Universidade de Sao Paulo/BR