Acurácia e precisão do método ultrassonográfico manual de aferição da espessura mediointimal carotídea tendo como referência o método semiautomático / Accuracy and precision of ultrasound manual measurement of carotid intima-media thickness having semiautomatic method as reference / Exactitud y Precisión del Método Ecográfico Manual de Medición del Espesor Mediointimal Carotideo, teniendo como referencia el Método Semiautomático
Garcia, Maristela Magnavita Oliveira; Ladeira, Ana Marice Teixeira; Lima, Isis Vasconcelos; Varela, Carolina Garcez; Silva, Patricia Fontes da Costa; Rodrigues, Marilia Galeffi; Silva, Maria de Lourdes Lima; Guimarães, Armênio Costa; Correia, Luis Claudio Lemos.
  • Garcia, Maristela Magnavita Oliveira; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Ladeira, Ana Marice Teixeira; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Lima, Isis Vasconcelos; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Varela, Carolina Garcez; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Silva, Patricia Fontes da Costa; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Rodrigues, Marilia Galeffi; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Silva, Maria de Lourdes Lima; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Guimarães, Armênio Costa; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
  • Correia, Luis Claudio Lemos; Fundação Bahiana para Desenvolvimento das Ciências. Escola Bahiana de Medicina e Saúde Pública. Salvador. BR
ABC., imagem cardiovasc ; 26(4): 276-283, out.-dez. 2013. ilus, tab, graf
Article in English, Spanish, Portuguese | LILACS | ID: lil-705123
RESUMO
Fundamento: A tecnologia semiautomática de detecção de bordas para aferição da Espessura Mediointimal Carotídea (EMIC), além de oferecer menor tempo de exame, apresenta mínima variabilidade das medidas.Objetivo: Testar a acurácia e a reprodutibilidade do método manual, tendo o semiautomático como referência.Métodos: Selecionadas mulheres do Ambulatório de Obesidade para aferição da EMIC. A avaliação do método manual (EMICmanual) foi realizada tomando-se a medida máxima do método semiautomático (EMICsamax) como referência, sendo utilizados teste de correlação de Pearson, grau de concordância e variabilidades inter e intraobservador nos dois modelos, além do teste de Kappa. Resultados: Na amostra de 59 participantes, houve forte correlação da medida de EMICmanual com EMICsamax (r = 0,84, p < 0,0001) e boa concordância entre ambos os métodos, dada a pequena média das diferenças de 0,06 ± 0,04 mm das medidas, com limites de concordância no nível 95 por cento entre -0,02 a 0,14 mm. A concordância na definição de aterosclerose subclínica foi moderada, de 53 por cento , (Kappa de 52 por cento p < 0,0001). Observou-se forte correlação com os métodos manual e semiautomático, entre as medidas intra e interobservador e boa reprodutibilidade intraobservador das medidas de EMICmanual, com diferença média de 0,04 ± 0,03 mm e limites de concordância no nível 95 por cento entre -0,02 a 0,10 mm, semelhantemente às medidas de EMICsamax, que se repetiu na reprodutibilidade interobservador. Conclusão: O método manual de aferição da EMIC pôde ser validado como método alternativo ao semiautomático.
ABSTRACT
Background: The semiautomated ultrasound border detection technology for measurement carotid intima-media thickness (CIMT) provides shorter time of exam and minimal variability measures.Objective: To evaluate the accuracy of the manual method, taking the semiautomated as reference and the reproducibility of the methods.Methods: Data were obtained from women of Ambulatory of Obesity directed to CIMT measurements determination. The evaluation of the manual method (manualCIMT) was performed, taking the maximum thickness of the semiautomated method (samaxCIMT) as reference. The statistical analyses used Pearson correlation, degree of agreement and inter and intraobserver variabilities in the two models and Kappa test.Results: In the sample of 59 women, there was a strong correlation of manualCIMT with samaxCIMT (r = 0.84, p <0.0001) and a good agreement between both methods, given the small mean difference of measures (0.06 ± 0.04 mm), with limits of agreement at 95 percent level between -0.02 to 0.14 mm. The agreement on the definition of subclinical atherosclerosis was moderate, 53 percent (kappa 52 percent, p <0.0001). The correlation between manual and semiautomated measurements in intra and inter-observer analyses was strong. The intraobserver reproducibility for manualCIMT measures was good and similar of interobserver, with a mean difference of 0.04 ± 0.03 mm and limits of agreement at 95 percent level between -0.02 to 0.10 mm. For EMICsamáx similar result was observed. Conclusion: The Manual method of measuring carotid intima-media thickness could be validated as na alternative to semi-automated.
RESUMEN
Justificación: La tecnología semiautomática de detección de bordes para medición del Espesor Mediointimal Carotideo (EMIC), además de ofrecer un menor tiempo de examen, presenta mínima variabilidad de las medidas. Objetivo: Poner a prueba la exactitud y la reproducibilidad del método manual, tomando el semiautomático como referencia. Métodos: Se han seleccionadas mujeres en la Clínica de Obesidad para medición de EMIC. La evaluación del método manual (EMICmanual) ha sido realizada tomándose la medida máxima del método semiautomático (EMICsamax) como referencia, ya que se utiliza prueba de correlación de Pearson, grado de concordancia y variabilidades inter e intraobservador en los dos modelos, además de la prueba de Kappa. Resultados: En la muestra de 59 participante, hubo fuerte correlación de la medida de EMICmanual con EMICsamax (r = 0,84, p < 0,0001) y una buena concordancia entre los dos métodos, dado el pequeño promedio de las diferencias de 0,06 ± 0,04 mm de las medidas, con límites de concordancia al nivel del 95% entre -0,02 a 0,14 mm. La concordancia en la definición de aterosclerosis subclínica ha sido moderada, del 53%, (Kappa del 52% p < 0,0001). Se ha observado una fuerte correlación con los métodos manual y semiautomático, entre las medidas intra e interobservador y buena reproducibilidad intraobservador de las medidas de EMICmanual, con diferencia media de 0,04 ± 0,03 mm y límites de concordancia al nivel del 95% entre -0,02 a 0,10 mm, de forma similar a las medidas de EMICsamax, que se repitió en la reproducibilidad inter-observador. Conclusión: El método manual de medición de EMIC puede evaluarse como método alternativo al semiautomático
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Ultrasonography / Benchmarking / Atherosclerosis / Carotid Intima-Media Thickness Type of study: Diagnostic study / Prognostic study Limits: Adult / Female / Humans Language: English / Spanish / Portuguese Journal: ABC., imagem cardiovasc Journal subject: Cardiology / Diagnostic Imaging Year: 2013 Type: Article Affiliation country: Brazil Institution/Affiliation country: Fundação Bahiana para Desenvolvimento das Ciências/BR

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Ultrasonography / Benchmarking / Atherosclerosis / Carotid Intima-Media Thickness Type of study: Diagnostic study / Prognostic study Limits: Adult / Female / Humans Language: English / Spanish / Portuguese Journal: ABC., imagem cardiovasc Journal subject: Cardiology / Diagnostic Imaging Year: 2013 Type: Article Affiliation country: Brazil Institution/Affiliation country: Fundação Bahiana para Desenvolvimento das Ciências/BR