Evaluación de la eficacia de los parámetros del electroencefalograma cuantitativo en la medición del nivel de profundidad anestésico / Evaluation of the effectiveness of the quantitative electroencephalogram parameters in the measurement of the anesthetic depth level
Montoya Pedrón, Arquímedes; Marañón Reyes, Enrique Juan; Rodríguez Aldana, Yissel; Álvarez Ruffo, Carlos Manuel; Salgado Castillo, Antonio.
  • Montoya Pedrón, Arquímedes; Hospital General Docente Dr. Juan Bruno Zayas Alfonso. Santiago de Cuba. CU
  • Marañón Reyes, Enrique Juan; Centro de Estudio de Neurociencias, Procesamiento de Imágenes y Señales de la Universidad de Oriente, Santiago de Cuba, Cuba. Santiago de Cuba. CU
  • Rodríguez Aldana, Yissel; Centro de Estudio de Neurociencias, Procesamiento de Imágenes y Señales de la Universidad de Oriente, Santiago de Cuba, Cuba. Santiago de Cuba. CU
  • Álvarez Ruffo, Carlos Manuel; Hospital General Docente Dr. Juan Bruno Zayas Alfonso. Santiago de Cuba. CU
  • Salgado Castillo, Antonio; Universidad de Oriente, Santiago de Cuba, Cuba. Santiago de Cuba. CU
Medisan ; 18(3)mar. 2014. ilus, tab
Article in Spanish | LILACS, CUMED | ID: lil-709135
RESUMEN

Introducción:

los métodos de medición del nivel de profundidad del efecto hipnótico de los fármacos anestésicos, a partir del análisis cuantitativo del electroencefalograma, necesitan ser perfeccionados y optimizados para garantizar su aplicación eficiente en la práctica clínica.

Objetivo:

evaluar los efectos del nivel de profundidad anestésica y de la derivación del registro en los parámetros del electroencefalograma cuantitativo, para garantizar la selección de los parámetros óptimos en la clasificación del nivel de profundidad anestésica.

Métodos:

se estudió una muestra de 29 adultos con afecciones abdominales, tratados quirúrgicamente por vía endoscópica, bajo anestesia general. El registro electroencefalográfico se realizó mediante un montaje de 19 canales y el nivel de profundidad anestésica fue cuantificado clínicamente mediante una escala de 8 niveles. Igualmente, los parámetros del electroencefalograma cuantitativo fueron estimados mediante el sistema de análisis del equipo Medicid 5 de Neuronic.

Resultados:

el nivel de profundidad anestésica presentó un efecto significativo en los parámetros del electroencefalograma cuantitativo, en los modelos espectrales de banda ancha y estrecha. Entre los parámetros con mayor significación figuraron el poder absoluto delta, theta, el poder relativo theta y la frecuencia media theta, alpha y total; mientras que en los parámetros de banda estrecha se obtuvo un efecto significativo en todas las derivaciones, con una interacción significativa entre la topografía y el nivel de profundidad anestésica.

Conclusiones:

los parámetros del electroencefalograma cuantitativo pueden ser utilizados de forma eficaz en la predicción del nivel de profundidad anestésica, con una mayor resolución en los niveles de clasificación que los utilizados hasta el presente. Asimismo, se confirmó el efecto selectivo de los agentes hipnóticos en las diferentes áreas corticales.
ABSTRACT

Introduction:

the depth level measurement methods of the hypnotic effect of anesthetic drugs, from the quantitative analysis of the electroencephalogram, need to be corrected and optimized to guarantee their efficient application in the clinical practice.

Objective:

to evaluate the effects of the anesthetic depth level and of the record derivation in the parameters of the quantitative electroencephalogram, to guarantee the selection of optimal parameters in the classification of the anesthetic depth level.

Methods:

a sample of 29 adults with abdominal disorders, surgically treated through endoscopy, under general anaesthesia was studied. The electroencephalographic record was carried out by means of a 19 channels assembly and the level of anesthetic depth was clinically quantified by means of an 8 levels scale. Equally, the parameters of the quantitative electroencephalogram were calculated by means of the analysis system of the Medicid 5 equipment of Neuronic.

Results:

the level of anesthetic depth presented a significant effect in the parameters of the quantitative electroencephalogram, in the spectral models of broad and narrow band. Among the parameters with more significance there were the absolute power delta, theta, the relative power theta and the mean frequency theta, alpha and total; while in the narrow band parameters a significant effect was obtained in all derivations, with a significant interaction between the topography and the anesthetic depth level.

Conclusions:

the parameters of the quantitative electroencephalogram can be used in an effective way in the prediction of the anesthetic depth level, with a higher resolution in the classification levels than those used up to now. Also, the selective effect of the hypnotic agents was confirmed in the different cortical areas.
Subject(s)

Full text: Available Index: LILACS (Americas) Main subject: Electroencephalography / Hypnotics and Sedatives Type of study: Prognostic study Language: Spanish Journal: Medisan Journal subject: Medicine Year: 2014 Type: Article Affiliation country: Cuba Institution/Affiliation country: Centro de Estudio de Neurociencias, Procesamiento de Imágenes y Señales de la Universidad de Oriente, Santiago de Cuba, Cuba/CU / Hospital General Docente Dr. Juan Bruno Zayas Alfonso/CU

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: LILACS (Americas) Main subject: Electroencephalography / Hypnotics and Sedatives Type of study: Prognostic study Language: Spanish Journal: Medisan Journal subject: Medicine Year: 2014 Type: Article Affiliation country: Cuba Institution/Affiliation country: Centro de Estudio de Neurociencias, Procesamiento de Imágenes y Señales de la Universidad de Oriente, Santiago de Cuba, Cuba/CU / Hospital General Docente Dr. Juan Bruno Zayas Alfonso/CU